APPLICATION INSTRUMENT FOR STERNAL ZIPFIX
Report
- Report Number
- 2939274-2019-61696
- Event Type
- Malfunction
- Date Received
- October 22, 2019
- Report Date
- October 7, 2019
- Manufacturer
- WRIGHTS LANE SYNTHES USA PRODUCTS LLC
- Product Code
- JDQ
- UDI-DI
- 10887587010847
- PMA / PMN Number
- K110789
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. PART: 03.501.080-US. LOT: 9812895. MANUFACTURING SITE: HÄGENDORF. RELEASE TO WAREHOUSE DATE: 05. FEB. 2016. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE LOT NUMBER, AND NO NON-CONFORMANCES WERE IDENTIFIED. SERVICE EVALUATION: THE CUSTOMER REPORTED THE DEVICE WOULD NOT TENSION PROPERLY. THE ITEM PASSED TESTING PER THE INSPECTION SHEET AND WORKED WITHIN NORMAL PARAMETERS. THE COMPLAINED ISSUE WAS NOT ABLE TO BE REPRODUCED. THE CAUSE OF THE COMPLAINED ISSUE IS UNKNOWN. THE ITEM PASSED SYNTHES FINAL INSPECTION ON 29-OCT-2019 AND WILL BE RETURNED TO THE CUSTOMER UPON COMPLETION OF THE SERVICE AND REPAIR PROCESS. ATTACHED SERVICE RECORD COMPLETED. FINALIZED SERVICE RECORD WILL BE ARCHIVED IN DOCUMENT MANAGEMENT SYSTEM. THE EVALUATION WAS UNCONFIRMED. THE DEVICE WAS DEEMED SERVICEABLE AND WILL BE RETURNED TO THE CUSTOMER, NO DESIGN OR MANUFACTURING ISSUES WERE IDENTIFIED THEREFORE IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
REPORTER IS SYNTHES SALES CONSULTANT. THE DEVICE HAS BEEN RECEIVED, THE INVESTIGATION IS IN PROGRESS, NO CONCLUSION COULD BE DRAWN AT THE TIME OF FILING THIS REPORT. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
IT WAS REPORTED THAT ON (B)(6) 2019, THE APPLICATION INSTRUMENT FOR STERNAL ZIPFIX WILL NOT TENSION PROPERLY. THE ISSUE WAS OBSERVED DURING TESTING/CLEANING. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE (1) APPLICATION INSTRUMENT FOR STERNAL ZIPFIX. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1013289 | APPLICATION INSTRUMENT FOR STERNAL ZIPFIX | CERCLAGE FIXATION | JDQ | WRIGHTS LANE SYNTHES USA PRODUCTS LLC | 03.501.080 | 9812895 | 10887587010847 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |