FDA Adverse Event
Malfunction
Summary report: N
LIFEPAK CR PLUS DEFIBRILLATOR
MDR report key: 922043
·
Received October 1, 2007
Report
- Report Number
- 3015876-2007-00367
- Event Type
- Malfunction
- Date Received
- October 1, 2007
- Report Date
- September 6, 2007
- Manufacturer
- PHYSIO-CONTROL, INC.
- Product Code
- MKJ
- PMA / PMN Number
- K011144
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
PHYSIO-CONTROL, INC. EVALUATED THE DEVICE. THE ROOT CAUSE WAS DETERMINED TO BE DUE TO A FAILURE OF THE CAPACITOR C177 ON THE ANALOG PCB ASSEMBLY. AFTER REPLACING THE ANALOG PCB ASSEMBLY, PROPER OPERATION WAS CONFIRMED AND THE DEVICE WAS RETURNED TO THE CUSTOMER.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DEVICE WOULD NOT OPERATE ON DC POWER. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEPAK CR PLUS DEFIBRILLATOR | MKJ | PHYSIO-CONTROL, INC. | CR PLUS | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA YR |