FDA Adverse Event Malfunction Summary report: N

BBL CULTURESWAB PLUS AMIES GEL WITHOUT CHARCO

MDR report key: 921761 · Received March 28, 2007

Report

Report Number
1119779-2007-00005
Event Type
Malfunction
Date Received
March 28, 2007
Date of Event
March 2, 2007
Manufacturer
COPAN ITALIA
Product Code
LIO
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

RECORDS FOR THIS LOT OF MATERIAL WERE EXAMINED AND SHOWED NO ANOMALIES DURING MANUFACTURER OF THIS PRODUCT. A 100% VISUAL INSPECTION WAS CONDUCTED PRIOR TO BULK PACKAGING. RETENTION SAMPLES OF THIS LOT WERE VISUALLY EXAMINED FOR MISSING TIPS. SAMPLES OF THIS LOT WERE ALSO RETESTED USING A WETABILITY AND MANUAL PULL TEST TO ASSURE NO DETACHMENT OF THE SWABS FROM THE SHAFT. ALL RESULTS WERE ACCEPTABLE. BASED ON PRODUCT TRENDING AND INVESTIGATION RESULTS, THIS APPEARS TO BE AN ISOLATED INCIDENT.

Description of Event or Problem · 1

A CUSTOMER CALLED TO REPORT THAT WHILE CULTURING A PATIENT'S INCISION IN THE OPERATING ROOM (OR), THE OR NOTICED THAT ONE OF THE RAYON TIPS (THIS SWAB HAS TWO TIPS) WAS MISSING WHEN THEY WERE REPLACING THE SWAB BACK INTO THE TRANSPORT TUBE. THE SURGEONS CONDUCTED A THOROUGH SEARCH OF THE PATIENT'S INCISION AND FAILED TO FIND THE RAYON TOP AND THEY DO NOT BELIEVE THE TIP WAS LEFT IN THE INCISION. THE TIP WAS NEVER LOCATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BBL CULTURESWAB PLUS AMIES GEL WITHOUT CHARCO LIO LIO COPAN ITALIA NA 8K7200

Patients

Seq Age Sex Outcome Treatment
1 * Other