CLINITEK STATUS+
Report
- Report Number
- 3002637618-2019-00112
- Event Type
- Malfunction
- Date Received
- October 21, 2019
- Date of Event
- September 25, 2019
- Report Date
- December 9, 2019
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS
- Product Code
- JHI
- PMA / PMN Number
- K091216
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- EI
- Reporter Occupation
- NURSE
Narratives
CORRECTION: CHANGED PRODUCT CODE - PRODUCT CODE FROM JIL TO JHI (USED FOR HCG). SIEMENS SERVICE REPORT: ALL TESTS WERE PERFORMED USING THE POWER SUPPLY THAT CAME IN WITH THE INSTRUMENT AND THE INSTRUMENT WAS CONNECTED TO ITS CONNECT PLATFORM WITH THE BARCODE READER PLUGGED IN. WHEN SERVICE UNPACKED THE TABLE ASSEMBLY, THEY NOTICED A MARK ON THE CALIBRATION BAR WHICH LOOKED LIKE DRY SPILLAGE. ON CLOSER INSPECTION WHEN THE MARKS WOULD NOT CLEAN OFF, THEY FOUND THAT THE MARKS WERE PHYSICAL DAMAGE TO THE BAR. THE TABLE ASSEMBLY WAS REPLACED AND TEN CASSETTES WERE RUN WITH NO FALSE RESULTS.
THE CUSTOMER STATED THAT REPEAT TESTING WAS PERFORMED TO CONFIRM CORRECT RESULTS WHICH MET THE CLINICAL PICTURE AND A CORRECTED REPORT WAS ISSUED. THE INSTRUMENT HAS BEEN REQUESTED FOR RETURN TO THE WORKSHOP FOR INVESTIGATION. PROPER TECHNIQUE AND MAINTENANCE WERE REVIEWED WITH THE CUSTOMER. THERE WERE NO ISSUES FOUND. THE CAUSE OF THIS EVENT IS UNKNOWN.
THE CUSTOMER REPORTED THE CLINITEK STATUS+ WAS GENERATING BORDERLINE HCG RESULTS ON NEGATIVE SAMPLES. ONE PATIENT'S PROCEDURE (UNKNOWN) WAS DELAYED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1007088 | CLINITEK STATUS+ | CLINITEK STATUS+ | JHI | SIEMENS HEALTHCARE DIAGNOSTICS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |