FDA Adverse Event Malfunction Summary report: N

NUCLETRON NPS

MDR report key: 92129 · Received May 16, 1997

Report

Report Number
1121753-1997-00021
Event Type
Malfunction
Date Received
May 16, 1997
Date of Event
January 25, 1997
Report Date
May 14, 1997
Manufacturer
NUCLETRON CORP.
Product Code
IWB
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

THE DECAY FACTOR FOR THE SOURCE IS NOT TAKEN INTO ACCOUNT WHEN RECALLING AN EXISTING PATIENT FILE AND WHEN NO TREATMENT PARAMETERS ARE CHANGED;I.E., WHEN RECALLING AN EXISTING FILE THE ORIGINAL SOURCE STRENGTH AND PARAMETERS ARE DISPLAYED AND ARE USED FOR TREATMENT PURPOSES. NOTE: THIS PROBLEM WAS PREVIOUSLY REPORTED AS 1121753-1997-0005.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NUCLETRON NPS RADIOTHERAPY TREATMENT PLANNING SYSTEM IWB NUCLETRON CORP. NPS BPS - MPS *

Patients

Seq Age Sex Outcome Treatment
1 * Other