FDA Adverse Event
Malfunction
Summary report: N
NUCLETRON NPS
MDR report key: 92129
·
Received May 16, 1997
Report
- Report Number
- 1121753-1997-00021
- Event Type
- Malfunction
- Date Received
- May 16, 1997
- Date of Event
- January 25, 1997
- Report Date
- May 14, 1997
- Manufacturer
- NUCLETRON CORP.
- Product Code
- IWB
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
THE DECAY FACTOR FOR THE SOURCE IS NOT TAKEN INTO ACCOUNT WHEN RECALLING AN EXISTING PATIENT FILE AND WHEN NO TREATMENT PARAMETERS ARE CHANGED;I.E., WHEN RECALLING AN EXISTING FILE THE ORIGINAL SOURCE STRENGTH AND PARAMETERS ARE DISPLAYED AND ARE USED FOR TREATMENT PURPOSES. NOTE: THIS PROBLEM WAS PREVIOUSLY REPORTED AS 1121753-1997-0005.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | NUCLETRON NPS | RADIOTHERAPY TREATMENT PLANNING SYSTEM | IWB | NUCLETRON CORP. | NPS BPS - MPS | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * | Other |