FDA Adverse Event
Malfunction
Summary report: N
PLATO-RTS
MDR report key: 92113
·
Received May 16, 1997
Report
- Report Number
- 1121753-1997-00019
- Event Type
- Malfunction
- Date Received
- May 16, 1997
- Date of Event
- March 26, 1997
- Report Date
- May 7, 1997
- Manufacturer
- NUCLETRON CORP.
- Product Code
- IWB
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- NOT APPLICABLE
Narratives
Description of Event or Problem · 1
FOR SOFTWARE PLATO RTS V1.X, INCORRECT MONITOR UNITS ARE CALCULATED IN A MULTIPLE ELECTRON BEAM SET-UP, IN WHICH MANTLE TYPE SHIELDING IS INCLUDED WHEN REFERENCE AXIS ARE IN DIFFERENT LOCATIONS FOR EACH FIELD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PLATO-RTS | RADIOTHERAPY TREATMENT PLANNING SYSTEM | IWB | NUCLETRON CORP. | RTS V 1.5 | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * | Other |