FDA Adverse Event Malfunction Summary report: N

PLATO-RTS

MDR report key: 92113 · Received May 16, 1997

Report

Report Number
1121753-1997-00019
Event Type
Malfunction
Date Received
May 16, 1997
Date of Event
March 26, 1997
Report Date
May 7, 1997
Manufacturer
NUCLETRON CORP.
Product Code
IWB
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
NOT APPLICABLE

Narratives

Description of Event or Problem · 1

FOR SOFTWARE PLATO RTS V1.X, INCORRECT MONITOR UNITS ARE CALCULATED IN A MULTIPLE ELECTRON BEAM SET-UP, IN WHICH MANTLE TYPE SHIELDING IS INCLUDED WHEN REFERENCE AXIS ARE IN DIFFERENT LOCATIONS FOR EACH FIELD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PLATO-RTS RADIOTHERAPY TREATMENT PLANNING SYSTEM IWB NUCLETRON CORP. RTS V 1.5 *

Patients

Seq Age Sex Outcome Treatment
1 * Other