FDA Adverse Event Injury Summary report: N

MENTOR SMOOTH ROUND MODERATE PROFILE

MDR report key: 9208217 · Received October 18, 2019

Report

Report Number
1645337-2019-22045
Event Type
Injury
Date Received
October 18, 2019
Date of Event
September 14, 2019
Report Date
September 19, 2019
Manufacturer
MENTOR TEXAS
Product Code
FWM
UDI-DI
00081317001287
PMA / PMN Number
P990075
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE INVESTIGATION OF THE RETURNED DEVICE WAS COMPLETED BY THE FAILURE ANALYSIS LAB ON 11/8/2019. DEVICE EVALUATION SUMMARY: ACCORDING TO THE INFORMATION PROVIDED, THE PATIENT EXPERIENCED A DEFLATION ON THE IMPLANT. DURING EVALUATION OF THE DEVICE A CREASE WAS NOTED ON THE POSTERIOR ASPECT. ALSO A TEAR WAS OBSERVED WITHIN THE CREASE MEASURING APPROXIMATELY 0.1 CM. NO OTHER ANOMALIES WERE OBSERVED. THESE KIND OF FINDINGS ARE CONSISTENT WITH A CREASE/FOLD FAILURE WHICH IS A KNOWN INHERENT RISK ASSOCIATED WITH THE USE OF SALINE-FILLED MAMMARY PROSTHESES AND MAY BE THE RESULT OF ONE OR MORE OF THE FOLLOWING CONTRIBUTING FACTORS: UNDERINFLATION OR OVERINFLATION OF THE DEVICE, CONTINUOUS AND SUSTAINED STRESSES TO THE DEVICE - SUCH A TOO SMALL A BREAST POCKET AND FOLDING OR WRINKLING OF THE SHELL IN THE BREAST POCKET. MANUFACTURER'S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 1

THE MENTOR FAILURE ANALYSIS LAB HAS RECEIVED THE DEVICE FOR EVALUATION. THE ANALYSIS HAS BEGUN BUT IS NOT COMPLETE AT THIS TIME. WHEN THE INVESTIGATIONAL ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: DEFLATION. CONCOMITANT PRODUCTS: 425CC MENTOR SMOOTH ROUND MODERATE PROFILE SALINE BREAST IMPLANT CATALOG: 3501670 LOT: 5841538 SERIAL NUMBER: (B)(4). MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT A (B)(6) YEAR OLD CAUCASIAN FEMALE PATIENT WHO UNDERWENT BREAST AUGMENTATION REVISION SURGERY WITH A 425CC MENTOR SMOOTH ROUND MODERATE PROFILE SALINE BREAST IMPLANT EXPERIENCED RIGHT SIDED DEFLATION POST PROCEDURE. AS A RESULT, PATIENT HAD UNDERGONE BILATERAL REMOVAL AND REPLACEMENT WITH A 525CC MENTOR MEMORYGEL BREAST IMPLANT ON (B)(6) 2019.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1000921 MENTOR SMOOTH ROUND MODERATE PROFILE PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE FWM MENTOR TEXAS 6662421 00081317001287

Patients

Seq Age Sex Outcome Treatment
1 40 YR Required Intervention