HISTOACRYL BLUE 0.5ML
Report
- Report Number
- 3003639970-2019-00716
- Event Type
- Malfunction
- Date Received
- October 17, 2019
- Date of Event
- September 19, 2019
- Report Date
- December 15, 2019
- Manufacturer
- B.BRAUN SURGICAL SA
- Product Code
- MPN
- PMA / PMN Number
- SEE H10
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NZ
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
SAMPLES RECEIVED: 1 OPEN POUCH. ANALYSIS AND RESULTS: THERE ARE NO PREVIOUS COMPLAINTS OF THIS CODE BATCH OF WHICH WE MANUFACTURED AND DISTRIBUTED IN THE MARKET (B)(4) OF THIS CODE-BATCH. THERE ARE NO UNITS IN B. BRAUN SURGICAL'S WAREHOUSE. WE HAVE RECEIVED ONE OPEN POUCH THAT CONTAINS AN UNOPENED AMPOULE. THE AMPOULE HAS BEEN OPTICALLY EVALUATED AND A DEFECT IN THE SEALING BAR OF THE AMPOULE (YELLOW BAR AT THE BOTTOM OF THE AMPOULE) WAS FOUND. THE LEAKAGE OF THE GLUE OCCURS AT THIS POINT. WE HAVE CONDUCTED A REVIEW OF THE BATCH MANUFACTURING RECORD OF THIS PRODUCT AND NO DEVIATIONS RELATED TO THIS ISSUE HAVE BEEN FOUND. FINAL CONCLUSION: TAKING INTO ACCOUNT THAT THE RESULTS OF THE SAMPLE RECEIVED DO NOT FULFIL B. BRAUN SURGICAL SPECIFICATIONS, WE CONCLUDE THAT THE COMPLAINT IS CONFIRMED BY EVIDENCE OF FAILURE IN THE SAMPLE RECEIVED. ACTIONS ON DISTRIBUTED PRODUCT OF THIS REFERENCE/BATCH: BASED ON THE CONCLUSION DERIVED FROM INVESTIGATION, IT IS NOT REQUIRED TO MAKE ACTIONS IN DISTRIBUTED PRODUCT. A CAPA HAS BEEN INITIATED.
PMA/510K: REPORTED DEVICE NOT MARKETED IN THE U.S., HOWEVER, SIMILAR DEVICES OR DEVICES THAT SHARE COMPONENTS, RAW MATERIALS, PROCESS METHODS OR OTHER TECHNOLOGICAL CHARACTERISTICS ARE REGISTERED WITHIN THE U.S. K111959. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, A FOLLOW UP REPORT WILL BE SUBMITTED.
THE REPORTER INDICATED THAT THE HISTOACRYL AMPULE HAD SHATTERED IN THE PACKAGING. ADDITIONAL INFORMATION WAS NOT AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 997365 | HISTOACRYL BLUE 0.5ML | TISSUE ADHESIVES | MPN | B.BRAUN SURGICAL SA | 1050044 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |