HEARTMATE 3 VAD MODULAR CABLE, INT
Report
- Report Number
- 2916596-2019-04700
- Event Type
- Injury
- Date Received
- October 17, 2019
- Date of Event
- September 29, 2019
- Report Date
- December 5, 2019
- Manufacturer
- THORATEC CORPORATION
- Product Code
- DSQ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SN
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
MANUFACTURER'S INVESTIGATION CONCLUSION: EVALUATION OF THE SUBMITTED PHOTO CONFIRMED DAMAGE ON THE PATIENT¿S MODULAR CABLE, LOT NUMBER 162759, TO THE INLINE CONNECTOR BEND RELIEF, AS WELL AS DISCOLORATION OF THE MODULAR CABLE, INLINE CONNECTOR BEND RELIEF, AND CONTROLLER CONNECTOR BEND RELIEF. THE ACCOUNT REPORTED THAT THE PATIENT¿S MODULAR CABLE WAS DAMAGED. THE DAMAGE WAS REPAIRED WITH RESCUE TAPE TEMPORARILY, BUT THE MODULAR CABLE WAS EXCHANGED ON 15 OCT 2019. THE DAMAGED MODULAR CABLE WAS NOT RETURNED FOR EVALUATION. THE HEARTMATE 3 LVAS IFU AND PATIENT HANDBOOK CONTAIN INFORMATION ON HOW TO CLEAN AND CARE FOR THE DRIVELINE. THE HM3 LVAS IFU ALSO EXPLAINS HOW TO REPLACE THE MODULAR CABLE. NO FURTHER INFORMATION WAS PROVIDED. THE MANUFACTURER IS CLOSING THE FILE ON THIS EVENT.
NO FURTHER INFORMATION WAS PROVIDED. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER¿S INVESTIGATION IS COMPLETED.
IT WAS REPORTED THAT THE PATIENT'S EXTERNAL PERCUTANEOUS LEAD HAD BEEN REPAIRED BY THE APPLICATION AND REINFORCEMENT OF RESCUE TAPE ON MULTIPLE OCCASIONS. THE LVAD PUMP IS FUNCTIONING AS INTENDED WITH NO ABNORMAL ALARMS DETECTED. UPON INSPECTION ON 29SEP2019, THE EXTERNAL SILICONE HOUSING NEAR THE MODULE CABLE CONNECTOR WAS ALMOST DETACHED. RESCUE TAPE WAS RE-APPLIED. THE MODULAR CABLE WAS EXCHANGED ON (B)(6) 2019. THE PATIENT WAS ASYMPTOMATIC WITH NO ISSUES DURING EXCHANGE AND THE MODULAR CABLE WILL NOT BE RETURNED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 996055 | HEARTMATE 3 VAD MODULAR CABLE, INT | VENTRICULAR (ASSISST) BYPASS | DSQ | THORATEC CORPORATION | 106525INT | 162759 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 33 YR |