FDA Adverse Event Injury Summary report: N

STERILIZER 16" CENTURY VAC

MDR report key: 919939 · Received September 28, 2007

Report

Report Number
2515984-2007-00001
Event Type
Injury
Date Received
September 28, 2007
Date of Event
September 4, 2007
Report Date
September 26, 2007
Manufacturer
STERIS CORPORATION-ERIE
Product Code
FLE
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

STERIS SVC TECH INSPECTED THE CENTURY STEAM STERILIZER (16" PREVAC). THE TECH DISASSEMBLED AND INSPECTED THE S-2 VALVE AND FOUND IT TO BE IN GOOD CONDITION. THE TECHNICIAN TESTED THE STERILIZER OPERATION IN BOTH SERVICE MODE AND OPERATOR MODE WITH NO PROBLEMS. THE TECHNICIAN WAS UNABLE TO DUPLICATE THE STERILIZER MALFUNCTION AS DESCRIBED BY THE INJURED OPERATOR. THE CENTURY SMALL OPERATOR'S MANUAL (129367-408) IN THE SUMMARY OF WARNINGS AND CAUTIONS (PAGES 1-1 THROUGH 1-3), STERILIZATION TECHNIQUES - BIOLOGICAL MONITORS (PAGES 2-4) AND STERILIZER OPERATION (PAGES 4-1,PAGES 4-3 THROUGH 4-4) CLEARLY INFORM POTENTIAL USERS OF THE PROPER INSTRUCTIONS FOR PLACING A BIOLOGICAL INDICATOR IN THE STERILIZER AND THE NEED FOR PERSONAL PROTECTION EQUIPMENT WHEN PLACING ITEMS WITHIN THE STERILIZER. STERIS COMPLAINT HISTORY OVER A FIVE YEAR PERIOD INDICATES NO OTHER CUSTOMER COMPLAINTS FOR THIS FACILITY.

Description of Event or Problem · 1

THE OPERATOR WAS LOADING A BIOLOGICAL INDICATOR INTO A CENTURY STEAM STERILIZER (16" PREVAC). THE OPERATOR WITH HER BARE HANDS WAS TRYING TO PLACE THE INDICATOR DIRECTLY ON THE RACK INSIDE THE STERILIZATION CHAMBER BESIDE/BETWEEN THE SHELVES. THE OPERATOR CLAIMS THAT A BURST OF STEAM ENTERED THE CHAMBER AND BURNED HER HAND. THERE WERE NO WITNESSES TO THIS INCIDENT. THE INJURED OPERATOR WAS SEEN IN THE HOSPITAL'S EMERGENCY ROOM AND TRANSFERRED TO THE HOSPITAL'S BURN UNIT. SKIN GRAFT SURGERY FOR THE OPERATOR WAS SCHEDULED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STERILIZER 16" CENTURY VAC STERILIZER FLE STERIS CORPORATION-ERIE 1021100 NA

Patients

Seq Age Sex Outcome Treatment
1 YR Required Intervention