TRUE METRIX
Report
- Report Number
- 1000113657-2019-10060
- Event Type
- Malfunction
- Date Received
- October 14, 2019
- Date of Event
- September 16, 2019
- Report Date
- October 30, 2019
- Manufacturer
- TRIVIDIA HEALTH INC
- Product Code
- NBW
- UDI-DI
- 00021292007836
- PMA / PMN Number
- K150052
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY, US
- Reporter Occupation
- OTHER
Narratives
METER WAS RETURNED FOR EVALUATION. NO DEFECT WAS DETECTED. TEST STRIPS WERE RETURNED FOR EVALUATION. DEFECT WAS DETECTED. MOST LIKELY UNDERLYING ROOT CAUSE: RC-61: STORAGE OUTSIDE SPECIFICATIONS. NOTE: MANUFACTURER CONTACTED CUSTOMER IN A FOLLOW-UP CALL TO ENSURE THE REPLACEMENT PRODUCTS RESOLVED THE INITIAL CONCERN - UNABLE TO ESTABLISH CONTACT WITH CUSTOMER AT THIS TIME. PRODUCT NOTIFICATION LETTER SENT TO CONTACT CUSTOMER CARE.
CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TEST RESULTS FROM RESULTS OBTAINED OF 233, 251, 200, 179 AND 172 MG/DL. THE CUSTOMER¿S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 100 - 130 MG/DL. CUSTOMER STATED THE HIGH RESULTS STARTED ABOUT ONE WEEK AGO. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL, A BACK TO BACK BLOOD TEST WAS NOT PERFORMED BY THE CUSTOMER; CUSTOMER HAD JUST TESTED PRIOR TO CALL AND HAD OBTAINED RESULT OF 233 MG/DL FASTING. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION IN THE LIVING AREA. THE TEST STRIP LOT MANUFACTURER¿S EXPIRATION DATE IS 10/31/2020 AND OPEN VIAL DATE IS 09/16/2019. CUSTOMER HAD BEEN USING TEST STRIP LOT NUMBER MV3197, BUT HAD FINISHED THE VIAL OF STRIPS. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: (B)(6).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 978070 | TRUE METRIX | SYSTEM, TEST BLOOD GLUCOSE, OVER THE COUNTER | NBW | TRIVIDIA HEALTH INC | STRIP, TRUE METRIX NFRS24/CS 50CT MG/DL | MW3301S | 00021292007836 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |