FDA Adverse Event Injury Summary report: N

STEM: AMISTEM H HA COATED STD STEM SIZE 2

MDR report key: 9178288 · Received October 11, 2019

Report

Report Number
3005180920-2019-00857
Event Type
Injury
Date Received
October 11, 2019
Date of Event
September 11, 2019
Report Date
October 11, 2019
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LZO
UDI-DI
07630030804083
PMA / PMN Number
K093944
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

BATCH REVIEW PERFORMED ON 23 SEPTEMBER 2019. LOT 140747: 86 ITEMS MANUFACTURED AND RELEASED ON 30-APR-2014. EXPIRATION DATE: 2019-03-30. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, 86 ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. CLINICAL EVALUATION PERFORMED BY MEDICAL AFFAIRS MANAGER. LATE INFECTION IN CEMENTLESS THA, MORE THAN 5 YEARS AFTER PRIMARY OPERATION. INFECTION IS A KNOWN POSSIBLE ADVERSE EVENT FOLLOWING EVERY SURGERY, INCLUDING THA'S. TO DATE, THERE IS NO REASON TO SUSPECT THAT THE CAUSE MAY BE LINKED TO THE IMPLANTED DEVICES. ADDITIONAL IMPLANT INVOLVED: CUP: VERSAFITCUP CC TRIO 01.26.45.1150 ACETABULAR SHELL CC TRIO NO-HOLE Ø 50 (K122911) LOT. 140831. BATCH REVIEW PERFORMED ON 23 SEPTEMBER 2019. LOT 140831: 100 ITEMS MANUFACTURED AND RELEASED ON 17-APR-2014. EXPIRATION DATE: 2019-03-30. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, 99 ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT.

Description of Event or Problem · 1

LOOSENING DUE TO INFECTION OF AN AMISTEM H 2 STD AND VERSAFITCUP CC TRIO 50MM. STEM AND CUP WERE SUCCESSFULLY REVISED. MORE THAN 5 YEARS AFTER PRIMARY SURGERY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
976936 STEM: AMISTEM H HA COATED STD STEM SIZE 2 CEMENTLESS HIP STEM LZO MEDACTA INTERNATIONAL SA 140747 07630030804083

Patients

Seq Age Sex Outcome Treatment
1 63 YR Required Intervention