FDA Adverse Event Other Summary report: N

AW

MDR report key: 917766 · Received March 16, 2007

Report

Report Number
9611343-2007-00016
Event Type
Other
Date Received
March 16, 2007
Date of Event
January 31, 2006
Report Date
March 16, 2007
Manufacturer
GE MEDICAL SYSTEMS SCS
Product Code
KPQ
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
PHYSICIST

Narratives

Description of Event or Problem · 1

IT WAS REPORTED BY THIS SITE THAT INCORRECT RADIOTHERAPY TREATMENT OCCURRED BECAUSE OF USER MISINTERPRETATION OF OUTPUT DATA PROVIDED BY GE ADVANTAGE SIM. THE SHIFT COORDINATES ARE GIVEN IN MILLIMETERS IN THE REPORT GENERATED BY ADVANTAGE SIM, BUT THEY ARE REQUIRED TO BE ENTERED IN CENTIMETERS IN THE NON-GE VARIAN ACCELERATORS. THE DIRECTION OF THE COORDINATES MAY HAVE ALSO BEEN MISINTERPRETED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AW * KPQ GE MEDICAL SYSTEMS SCS ADVANTAGE SIM 6, MD NA

Patients

Seq Age Sex Outcome Treatment
1 NO INFO Other