FDA Adverse Event
Other
Summary report: N
AW
MDR report key: 917766
·
Received March 16, 2007
Report
- Report Number
- 9611343-2007-00016
- Event Type
- Other
- Date Received
- March 16, 2007
- Date of Event
- January 31, 2006
- Report Date
- March 16, 2007
- Manufacturer
- GE MEDICAL SYSTEMS SCS
- Product Code
- KPQ
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- PHYSICIST
Narratives
Description of Event or Problem · 1
IT WAS REPORTED BY THIS SITE THAT INCORRECT RADIOTHERAPY TREATMENT OCCURRED BECAUSE OF USER MISINTERPRETATION OF OUTPUT DATA PROVIDED BY GE ADVANTAGE SIM. THE SHIFT COORDINATES ARE GIVEN IN MILLIMETERS IN THE REPORT GENERATED BY ADVANTAGE SIM, BUT THEY ARE REQUIRED TO BE ENTERED IN CENTIMETERS IN THE NON-GE VARIAN ACCELERATORS. THE DIRECTION OF THE COORDINATES MAY HAVE ALSO BEEN MISINTERPRETED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AW | * | KPQ | GE MEDICAL SYSTEMS SCS | ADVANTAGE SIM 6, MD | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NO INFO | Other |