FDA Adverse Event Malfunction Summary report: N

BD PLASTIPAK 50ML LUER-LOK SYRINGE

MDR report key: 9174877 · Received October 10, 2019

Report

Report Number
3003152976-2019-00710
Event Type
Malfunction
Date Received
October 10, 2019
Date of Event
September 23, 2019
Report Date
November 14, 2019
Manufacturer
BECTON DICKINSON, S.A.
Product Code
FMF
PMA / PMN Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: ONE SEALED SAMPLE OF 50LL LOT 1905216 WAS PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. THROUGH VISUAL INSPECTION OF THE SAMPLE, NO DAMAGE WAS NOTED IN THE PLUNGER ROD OR OTHER COMPONENTS THAT COULD HAVE CAUSED A LEAK AND NO LEAKAGE WAS OBSERVED. TEN RETAINED SAMPLES OF 50LL LOT 1905240 WERE EVALUATED. ON VISUAL INSPECTION OF THESE TEN SAMPLES, NO DAMAGE OR MOLDING DEFECTS CAN BE OBSERVED AND THE STOPPER IS CORRECTLY ASSEMBLED. A LEAK TEST WAS ALSO PERFORMED ON THE TEN RETAINED SAMPLES WHICH MET SPECIFICATIONS. A DEVICE HISTORY RECORD REVIEW FOUND NO NON-CONFORMANCES ASSOCIATED WITH THIS ISSUE DURING PRODUCTION OF THE REPORTED LOTS. BASED UPON THE VISUAL INSPECTION OF THE RETAINED SAMPLES, THE SAMPLE RECEIVED AND THE QUALITY TEAM'S INVESTIGATION, WE WERE UNABLE TO DETERMINE THE ROOT CAUSE FOR THIS INCIDENT.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD PLASTIPAK¿ 50ML LUER-LOK SYRINGE LEAKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: LIQUID RUNS OUT ON THE SYRINGE PLUNGER.

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT BD PLASTIPAK¿ 50ML LUER-LOK SYRINGE LEAKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: LIQUID RUNS OUT ON THE SYRINGE PLUNGER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
968344 BD PLASTIPAK 50ML LUER-LOK SYRINGE SYRINGE FMF BECTON DICKINSON, S.A. 1905216

Patients

Seq Age Sex Outcome Treatment
1 Other