BD MICROTAINER® TUBES WITH K2E (K2EDTA)
Report
- Report Number
- 2618282-2019-00257
- Event Type
- Malfunction
- Date Received
- October 9, 2019
- Date of Event
- September 23, 2019
- Report Date
- October 24, 2019
- Manufacturer
- BD CARIBE LTD.
- Product Code
- JKA
- UDI-DI
- 50382903659741
- PMA / PMN Number
- K991702
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
INVESTIGATION SUMMARY: BD HAD NOT RECEIVED SAMPLES OR PHOTOS FROM THE CUSTOMER FACILITY FOR EVALUATION. RETENTION SAMPLES WERE SELECTED FROM BD INVENTORY FOR EVALUATION/TESTING AND UPON COMPLETION, NO ISSUES WERE OBSERVED RELATING TO CLOTTING AS ALL SAMPLES MET SPECIFICATIONS. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR THE INCIDENT LOT AND, BASED ON THIS REVIEW, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET; THERE WERE NO RELATED QUALITY NON-CONFORMANCE'S DURING MANUFACTURING OF THE PRODUCT.
IT WAS REPORTED THAT BLOOD CLOTTED IN THE BD MICROTAINER® TUBES WITH K2E (K2EDTA) DURING USE. THIS OCCURRED ON 5 SEPARATE OCCASIONS, BUT THE DATES AND/OR PATIENT INFORMATION ARE UNKNOWN. THIS COMPLAINT WAS CREATED TO CAPTURE THE 1ST OF 2 RELATED INCIDENTS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "WE HAVE BEEN HAVING A LARGE AMOUNT OF CLOTTED CBC¿S IN THE LAV MICROTAINERS. WE CURRENTLY HAVE LOTS 9065638,9144756, AND 9193433. HAVE YOU BEEN HAVING ANY ISSUES? SOME OF THEM LOOK AS THOUGH THEY DONT HAVE ANY ADDITIVE IN THEM AT ALL."
A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT BLOOD CLOTTED IN THE BD MICROTAINER® TUBES WITH K2E (K2EDTA) DURING USE. THIS OCCURRED ON 5 SEPARATE OCCASIONS, BUT THE DATES AND/OR PATIENT INFORMATION ARE UNKNOWN. THIS COMPLAINT WAS CREATED TO CAPTURE THE 1ST OF 2 RELATED INCIDENTS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "WE HAVE BEEN HAVING A LARGE AMOUNT OF CLOTTED CBC¿S IN THE LAV MICROTAINERS. WE CURRENTLY HAVE LOTS 9065638,9144756, AND 9193433. HAVE YOU BEEN HAVING ANY ISSUES? SOME OF THEM LOOK AS THOUGH THEY DON¿T HAVE ANY ADDITIVE IN THEM AT ALL."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 964621 | BD MICROTAINER® TUBES WITH K2E (K2EDTA) | BLOOD SPECIMEN COLLECTION DEVICE | JKA | BD CARIBE LTD. | 9065638 | 50382903659741 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |