I-STAT EC4+ CARTRIDGE
Report
- Report Number
- 2245578-2019-00231
- Event Type
- Malfunction
- Date Received
- October 9, 2019
- Date of Event
- March 13, 2019
- Report Date
- December 4, 2019
- Manufacturer
- ABBOTT POINT OF CARE
- Product Code
- JGS
- UDI-DI
- 10054749000095
- PMA / PMN Number
- K912387
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
APOC INCIDENT # (B)(4). THE INVESTIGATION WAS COMPLETED ON 12/04/2019. A REVIEW OF THE DEVICE HISTORY RECORDS (DHRS) CONFIRMED THE CARTRIDGE LOTS MET FINISHED GOODS (FG) RELEASE CRITERIA. DUE TO THE EXPIRATION DATE OF EC4+ LOT K18306, RETAIN TESTING COULD NOT BE PERFORMED. RETAINED TESTING FOR EC4+ LOTS K18231 AND K19010 MET THE ACCEPTANCE CRITERIA OUTLINED IN APPENDIX 1 OF Q04.01.003 REV. AD (PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATION PROCEDURE). NO DEFICIENCY HAS BEEN DETERMINED FOR EC4+ LOTS K18231, K18306 AND K19010.
APOC INCIDENT (B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.
ON (B)(6) 2019, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT EC4+ CARTRIDGES THAT YIELDED A SUSPECTED DISCREPANT POTASSIUM RESULTS ON A (B)(6) YEAR OLD MALE PATIENT HAVING OUT-PATIENT SURGICAL PROCEDURE. THERE WAS NO ADDITIONAL PATIENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT. RETURN PRODUCT IS NOT AVAILABLE FOR INVESTIGATION. METHOD: I-STAT, DATE: 03/13/2019, TESTED: 07:13, RESULT: 7.2 MMOL/L. I-STAT, 03/13/2019, 07:21, 5.0 MMOL/L. COLLECTION TIME NOT PROVIDED. THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON LIMITED INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY. THE INVESTIGATION IS UNDERWAY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 967543 | I-STAT EC4+ CARTRIDGE | EC4+ CARTRIDGE | JGS | ABBOTT POINT OF CARE | NA | K18306 | 10054749000095 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 61 YR |