FDA Adverse Event Malfunction Summary report: N

I-STAT EC4+ CARTRIDGE

MDR report key: 9171497 · Received October 9, 2019

Report

Report Number
2245578-2019-00232
Event Type
Malfunction
Date Received
October 9, 2019
Date of Event
March 15, 2019
Report Date
December 4, 2019
Manufacturer
ABBOTT POINT OF CARE
Product Code
JGS
UDI-DI
10054749000095
PMA / PMN Number
K912387
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

APOC INCIDENT # (B)(4). THE INVESTIGATION WAS COMPLETED ON 12/04/2019. A REVIEW OF THE DEVICE HISTORY RECORDS (DHRS) CONFIRMED THE CARTRIDGE LOTS MET FINISHED GOODS (FG) RELEASE CRITERIA. DUE TO THE EXPIRATION DATE OF EC4+ LOT K18306, RETAIN TESTING COULD NOT BE PERFORMED. RETAINED TESTING FOR EC4+ LOTS K18231 AND K19010 MET THE ACCEPTANCE CRITERIA OUTLINED IN APPENDIX 1 OF Q04.01.003 REV. AD (PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATION PROCEDURE). NO DEFICIENCY HAS BEEN DETERMINED FOR EC4+ LOTS K18231, K18306 AND K19010.

Additional Manufacturer Narrative · 1

APOC INCIDENT # (B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.

Description of Event or Problem · 1

ON (B)(6) 2019, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT EC4+ CARTRIDGES THAT YIELDED A SUSPECTED DISCREPANT POTASSIUM RESULTS ON A (B)(6) YEAR OLD FEMALE PATIENT HAVING OUT-PATIENT SURGICAL PROCEDURE. THERE WAS NO ADDITIONAL PATIENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT. RETURN PRODUCT IS NOT AVAILABLE FOR INVESTIGATION. METHOD: I-STAT, DATE: (B)(6) 2019, TESTED: 11:34, RESULT: 6.9 MMOL/L. I-STAT, (B)(6) 2019. 11:43, 3.9 MMOL/L. COLLECTION TIME NOT PROVIDED. THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON LIMITED INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY. THE INVESTIGATION IS UNDERWAY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
967231 I-STAT EC4+ CARTRIDGE EC4+ CARTRIDGE JGS ABBOTT POINT OF CARE NA K18306 10054749000095

Patients

Seq Age Sex Outcome Treatment
1 75 YR