FDA Adverse Event Malfunction Summary report: N

OMNIPOD INSULIN PUMP

MDR report key: 9166127 · Received October 8, 2019

Report

Report Number
3004464228-2019-09868
Event Type
Malfunction
Date Received
October 8, 2019
Date of Event
September 25, 2019
Report Date
October 2, 2019
Manufacturer
INSULET CORPORATION
Product Code
LZG
UDI-DI
20385081120002
PMA / PMN Number
K122953
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED FOR EVALUATION. WE ARE UNABLE TO CONFIRM THE BENT CANNULA OR TO DETERMINE IF IT COULD HAVE CONTRIBUTED TO THE REPORTED HYPERGLYCEMIA. NO LOT RELEASE RECORDS WERE REVIEWED, AS THE PRODUCT LOT NUMBER WAS NOT PROVIDED. OMNIPOD INSULIN MANAGEMENT SYSTEM ¿ USER GUIDE: MODEL: UST400, 17845-5A-AW REV B 09/17. CHECKING YOUR BLOOD GLUCOSE: CHAPTER 4 / PAGE 36; WARNINGS: TEST RESULTS BELOW 70 MG/DL MEAN LOW BLOOD GLUCOSE (HYPOGLYCEMIA). TEST RESULTS GREATER THAN 250 MG/DL MEAN HIGH BLOOD GLUCOSE (HYPERGLYCEMIA). IF YOU GET RESULTS BELOW 70 MG/DL OR ABOVE 250 MG/DL, BUT DO NOT HAVE SYMPTOMS OF HYPOGLYCEMIA OR HYPERGLYCEMIA (SEE "LIVING WITH DIABETES" ON PAGE 115), REPEAT THE TEST. IF YOU HAVE SYMPTOMS OR CONTINUE TO GET RESULTS THAT FALL BELOW 70 MG/DL OR ABOVE 250 MG/DL, FOLLOW THE TREATMENT ADVICE OF YOUR HEALTHCARE PROVIDER.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT'S BLOOD GLUCOSE LEVELS REACHED 380 MG/DL WHILE WEARING THE POD BETWEEN 1 AND 4 HOURS. THE PATIENT REPORTS INSULIN WAS NOT BEING DELIVERED AND WHEN REMOVED FROM THE INFUSION SITE (ARM), THE POD'S CANNULA WAS FOUND BENT. AS TREATMENT, THE PATIENT WAS DELIVERED A MANUAL INJECTION OF INSULIN AND A NEW POD WAS APPLIED BRING THEIR BLOOD GLUCOSE LEVEL DOWN TO 98 MG/DL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
961829 OMNIPOD INSULIN PUMP PUMP, INFUSION, INSULIN LZG INSULET CORPORATION 14000 20385081120002

Patients

Seq Age Sex Outcome Treatment
1 40 YR