NOVOSYN VIOLET 0 (3.5) 70CM HRT40 (M)
Report
- Report Number
- 3003639970-2019-00708
- Event Type
- Injury
- Date Received
- October 7, 2019
- Report Date
- December 2, 2019
- Manufacturer
- B.BRAUN SURGICAL SA
- Product Code
- GAM
- PMA / PMN Number
- SEE H10
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
INVESTIGATION: SAMPLES RECEIVED: 22 UNOPENED POUCHES. ANALYSIS AND RESULTS: THERE ARE NO PREVIOUS COMPLAINTS OF THIS CODE-BATCH. WE MANUFACTURED AND DISTRIBUTED IN THE MARKET (B)(4) UNITS OF THIS CODE-BATCH. THERE ARE NO UNITS IN OUR STOCK. WE HAVE RECEIVED 22 CLOSED SAMPLES FOR ANALYSIS. TIGHTNESS TEST TO THE SAMPLES RECEIVED HAS BEEN PERFORMED AND THE UNITS ARE TIGHT. WE HAVE TESTED THE NEEDLE ATTACHMENT STRENGTH OF ALL SAMPLES RECEIVED AND THE RESULTS FULFIL THE REQUIREMENTS OF THE EUROPEAN PHARMACOPOEIA (EP): 1.77 KGF IN AVERAGE AND 1.16 KGF IN MINIMUM (EP REQUIREMENTS: 1.53 KGF IN AVERAGE AND 0.46 KGF IN MINIMUM). REVIEWED THE BATCH MANUFACTURING RECORD, THIS PRODUCT HAD A NORMAL PROCESS AND THE RESULTS DURING THE PROCESS FULFIL USP/EP AND B.BRAUN SURGICAL REQUIREMENTS. FINAL CONCLUSION: ALTHOUGH THE RESULTS OF THE SAMPLES RECEIVED FULFIL THE SPECIFICATIONS OF EUROPEAN PHARMACOPOEIA/ B. BRAUN SURGICAL SPECIFICATIONS, WE TAKE NOTE OF THIS INCIDENCE IN ORDER TO ASSESS IF NEW OR ADDITIONAL ACTIONS ARE NEEDED. ACTIONS ON PRODUCT DISTRIBUTED OF THIS REFERENCE/BATCH: BASED ON THE CONCLUSION DERIVED FROM INVESTIGATION, IT IS NOT REQUIRED TO MAKE ACTIONS IN DISTRIBUTED PRODUCT. NO CORRECTIVE/PREVENTIVE ACTIONS NEEDED.
PMA/510K: REPORTED DEVICE NOT MARKETED IN THE U.S., HOWEVER, SIMILAR DEVICES OR DEVICES THAT SHARE COMPONENTS, RAW MATERIALS, PROCESS METHODS OR OTHER TECHNOLOGICAL CHARACTERISTICS ARE REGISTERED WITHIN THE U.S. K122734. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, A FOLLOW UP REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT THE SUTURE DETACHES FROM THE NEEDLE. THE REPORTER INDICATED THAT THE THREAD DETACHED FROM NEEDLE WITHOUT USING MUSCULAR STRENGTH. IT HAS HAPPENED 4 TIMES WITH SUTURE FROM SAME BOX. NO FURTHER INFORMATION HAS BEEN RECEIVED. PATIENT INFORMATION IS NOT AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 958194 | NOVOSYN VIOLET 0 (3.5) 70CM HRT40 (M) | SYNTHETIC ABSORBABLE BRAIDED S | GAM | B.BRAUN SURGICAL SA | C0068024 | 719081 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |