FDA Adverse Event Malfunction Summary report: N

VIDAS® ESTRADIOL II

MDR report key: 9162981 · Received October 7, 2019

Report

Report Number
8020790-2019-00059
Event Type
Malfunction
Date Received
October 7, 2019
Report Date
January 23, 2020
Manufacturer
BIOMERIEUX SA
Product Code
CHP
PMA / PMN Number
K955647
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

AN INTERNAL INVESTIGATION WAS PERFORMED FOLLOWING A CUSTOMER REPORT OF FALSELY ELEVATED RESULTS FOR TWO FEMALE PATIENTS IN ASSOCIATION WITH THE VIDAS® ESTRADIOL II ASSAY (REF. 30431, LOT 1007168380). THE CUSTOMER REPORTED THE RESULTS WERE DISCREPANT BETWEEN VIDAS ESTRADIOL II AND RESULTS OBTAINED FROM AN EXTERNAL LABORATORY. BIOMÉRIEUX REQUESTED THE PATIENT SAMPLES, HOWEVER THEY WERE NOT SUBMITTED FOR THE INVESTIGATION. THE INVESTIGATION INCLUDED A REVIEW OF QUALITY CONTROL RECORDS AND INTERNAL SAMPLE CONTROL CHARTS, AS WELL AS TESTING ON INTERNAL AND EXTERNAL SAMPLES WITH VIDAS ESTRADIOL II LOT REF 30431 LOT 1007168380 / 200219-0. QUALITY CONTROL RECORDS: THE REVIEW OF VIDAS ESTRADIOL II BATCH HISTORY RECORDS FOR LOT 1007168380 / 200219-0 SHOWED NO ANOMALY DURING THE MANUFACTURING, CONTROL AND PACKAGING PROCESSES. THERE IS NEITHER CAPA NOR NON-CONFORMITY RECORDED ON VIDAS ESTRADIOL II TEST REFERENCE 30431 1007168380 / 200219-0 LINKED TO THE CUSTOMER'S ISSUE. STUDY OF INTERNAL SAMPLES CONTROL CHARTS: THIS ANALYSIS WAS CARRIED OUT : - ON FIVE (5) INTERNAL SAMPLES WITH TARGETS CLOSE TO THE RESULTS OBTAINED BY THE EXTERNAL LABORATORY. - ON SEVEN (7) BATCHES INCLUDING THE BATCH MENTIONED BY THE CUSTOMER COMPARED TO SIX OTHER BATCHES. ALL THE RESULTS ARE WITHIN SPECIFICATIONS AND IN THE TREND OF THE OTHER LOTS. INTERNAL TESTING PERFORMED: INTERNAL TESTING WAS PERFORMED ON INTERNAL SAMPLES WITH THE VIDAS ESTRADIOL II ASSAY, LOT 1007168380 / 200219-0 AND VIDAS ESTRADIOL II BATCH 1007566940/200520-0. THE COMPLAINTS LABORATORY TESTED FIVE (5) INTERNAL SAMPLES AND FOUR (4) FRESH BLOOD DONOR SAMPLES COLLECTED FROM WOMEN OVER THE AGE OF SIXTY. VIDAS ESTRADIOL II ASSAY, LOT 1007168380 / 200219-0 SHOW NO OVERESTIMATED RESULTS. CONCLUSION: THE INVESTIGATION SHOWED NO ANOMALIES DURING THE MANUFACTURING PROCESS, NO OVERESTIMATED RESULTS WERE OBTAINED DURING TESTING WITH FIVE (5) INTERNAL SAMPLES AND FOUR (4) FRESH BLOOD SAMPLES. BASED ON THIS DATA, THE KIT VIDAS ESTRADIOL II REF 30431 LOT 1007168380 / 200219-0 IS WITHIN THE EXPECTED PERFORMANCE PARAMETERS. THE VIDAS ESTRADIOL II INSTRUCTIONS FOR USE (IFU) "LIMITATIONS OF THE METHOD" SECTION STATES: DO NOT USE THE VIDAS® ESTRADIOL II ASSAY TO MEASURE ESTRADIOL LEVELS IN PATIENTS UNDERGOING FULVESTRANT THERAPY. INTERFERENCE MAY BE ENCOUNTERED WITH CERTAIN SERA CONTAINING ANTIBODIES DIRECTED AGAINST REAGENT COMPONENTS. FOR THIS REASON, ASSAY RESULTS SHOULD BE INTERPRETED TAKING INTO CONSIDERATION THE PATIENT'S HISTORY, AND THE RESULTS OF ANY OTHER TESTS PERFORMED.

Description of Event or Problem · 1

A CUSTOMER IN (B)(6) NOTIFIED BIOMÉRIEUX OF FALSELY OVER-ESTIMATED RESULTS IN ASSOCIATION WITH THE VIDAS® ESTRADIOL II ASSAY (REF. 30431, LOT 1007168380) PROCESSED ON A VIDAS 3 INSTRUMENT. TEST RESULTS ((B)(6) YEAR-OLD FEMALE): VIDAS ESTRADIOL II RESULT - 187 PG/ML. REPEAT VIDAS ESTRADIOL II TEST WAS NOT PERFORMED. EXTERNAL LABORATORY RESULT - 70 PG/ML. PATIENT IS MENOPAUSAL RECEIVING HORMONE REPLACEMENT THERAPY. THERE IS NO INDICATION OR REPORT FROM THE LABORATORY THAT THE DISCREPANT RESULT LED TO ANY ADVERSE EVENT RELATED TO THE PATIENT'S STATE OF HEALTH. THE PATIENT SAMPLE WAS REQUESTED FOR SUBMISSION. BIOMÉRIEUX INVESTIGATION WILL BE INITIATED. NOTE: ALTHOUGH PRODUCT REFERENCE 30431 IS NOT SOLD OR DISTRIBUTED IN THE UNITED STATES, A SIMILAR PRODUCT (REF. 30431-01) IS REGISTERED/SOLD IN THE UNITED STATES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
958151 VIDAS® ESTRADIOL II VIDAS® ESTRADIOL II CHP BIOMERIEUX SA 1007168380

Patients

Seq Age Sex Outcome Treatment
1 48 YR