FDA Adverse Event Malfunction Summary report: N

BD CONNECTA¿ STOPCOCK

MDR report key: 9161882 · Received October 7, 2019

Report

Report Number
9610847-2019-00606
Event Type
Malfunction
Date Received
October 7, 2019
Date of Event
September 20, 2019
Report Date
November 21, 2019
Manufacturer
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
Product Code
FMG
PMA / PMN Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FI
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 9122560. OUR RECORDS SHOW THAT THIS IS THE ONLY INSTANCE OF THIS ISSUE OCCURRING IN THIS PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. ADDITIONALLY, OUR ENGINEERS EVALUATED THE RETURNED SAMPLES, BUT WERE UNABLE TO OBSERVE ANY NON-CONFORMANCE THAT COULD HAVE LEAD TO A LOOSENED CAP. OUR ENGINEERS WERE ABLE TO TIGHTEN AND REMOVE THE CAP WITHOUT EXCESS EFFORT. THE ROOT CAUSE FOR THIS COMPLAINT COULD NOT BE DETERMINED AT THE CONCLUSION OF OUR REVIEW.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHEN OPENING THE PACKAGE THE CAP OF CONNECTA FALLS OFF WITH A BD CONNECTA¿ STOPCOCK. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: WHEN OPENING THE PRODUCT PACKAGE CAP OF CONNECTA GETS OFF TOO EASILY. IT DROPS OFF TO FLOOR AND GETS CONTAMINATED. CUSTOMER IS REQUESTING FACTORY WOULD FIX CAP TIGHTER.

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT WHEN OPENING THE PACKAGE THE CAP OF CONNECTA FALLS OFF WITH A BD CONNECTA¿ STOPCOCK. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: WHEN OPENING THE PRODUCT PACKAGE CAP OF CONNECTA GETS OFF TOO EASILY. IT DROPS OFF TO FLOOR AND GETS CONTAMINATED. CUSTOMER IS REQUESTING FACTORY WOULD FIX CAP TIGHTER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
958758 BD CONNECTA¿ STOPCOCK STOPCOCK FMG BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. 9122560

Patients

Seq Age Sex Outcome Treatment
1 Other