FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® SST¿ BLOOD COLLECTION TUBES

MDR report key: 9161514 · Received October 7, 2019

Report

Report Number
1024879-2019-01757
Event Type
Malfunction
Date Received
October 7, 2019
Date of Event
September 20, 2019
Report Date
October 8, 2019
Manufacturer
BECTON, DICKINSON & CO., (BD)
Product Code
JKA
UDI-DI
50382903679862
PMA / PMN Number
BK050036
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION: BD HAD NOT RECEIVED SAMPLES, BUT PHOTOS WERE PROVIDED BY THE CUSTOMER FACILITY FOR INVESTIGATION. THE PHOTOS WERE EVALUATED AND THE CUSTOMER'S INDICATED FAILURE MODE FOR LABEL LIFT WITH THE INCIDENT LOT WAS OBSERVED. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR THE INCIDENT LOT AND, BASED ON THIS REVIEW, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET; THERE WERE NO RELATED QUALITY NON-CONFORMANCES DURING MANUFACTURING OF THE PRODUCT. BD HAS INITIATED FURTHER INVESTIGATION RELATING TO THIS ISSUE THROUGH CAPA #1064141. THE INVESTIGATION IS STILL ON-GOING AND IMPROVEMENTS WILL BE MADE AS THE POTENTIAL CAUSES OF THIS ISSUE ARE IDENTIFIED.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD VACUTAINER® SST¿ BLOOD COLLECTION TUBES LABELS ARE LIFTING OFF. THIS WAS DISCOVERED BEFORE USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO. 367986, BATCH NO. 9093690 , 9030857. IT WAS REPORTED THAT LABELS ARE LIFTING FROM THE TUBES. PER EMAIL: QUANTITY : 9 BOXES. DETAILS OF THE COMPLAINT: LABELS COMES OFF.

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. MULTIPLE LOT NUMBERS: THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: DMEDICAL DEVICE LOT #: 9093690. MEDICAL DEVICE EXPIRATION DATE: 2020-03-31. DEVICE MANUFACTURE DATE: 2019-04-03. MEDICAL DEVICE LOT #: 9030857. MEDICAL DEVICE EXPIRATION DATE: 2020-01-31. DEVICE MANUFACTURE DATE: 2019-01-30. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT BD VACUTAINER® SST¿ BLOOD COLLECTION TUBES LABELS ARE LIFTING OFF. THIS WAS DISCOVERED BEFORE USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO. 367986. BATCH NO. 9093690 , 9030857. IT WAS REPORTED THAT LABELS ARE LIFTING FROM THE TUBES. PER EMAIL: QUANTITY : 9 BOXES. DETAILS OF THE COMPLAINT LABELS COMES OFF.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
958750 BD VACUTAINER® SST¿ BLOOD COLLECTION TUBES BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON & CO., (BD) SEE H.10 50382903679862

Patients

Seq Age Sex Outcome Treatment
1 Other