UNKNOWN
Report
- Report Number
- 3002808486-2019-01658
- Event Type
- Injury
- Date Received
- October 7, 2019
- Report Date
- December 10, 2019
- Manufacturer
- WILLIAM COOK EUROPE
- Product Code
- DTK
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER
Narratives
ADDITIONAL INFORMATION PROVIDED DETERMINED THAT THIS DEVICE WAS MANUFACTURED BY COOK INC. WITH THE SUBMISSION OF THIS FOLLOW UP REPORT, WILLIAM COOK EUROPE INFORMS THAT THIS COMPLAINT HAS BEEN TRANSFERRED FROM WILLIAM COOK EUROPE TO COOK INC. THIS REPORT IS REQUIRED BY THE FDA UNDER 21 CFR PART 803. THIS REPORT IS BASED ON UNCONFIRMED INFORMATION SUBMITTED BY OTHERS. NEITHER THE SUBMISSION OF THIS REPORT NOR ANY STATEMENT MADE IN IT IS INTENDED TO BE AN ADMISSION THAT ANY COOK DEVICE IS DEFECTIVE OR MALFUNCTIONED; THAT A DEATH OR SERIOUS INJURY OCCURRED; OR THAT ANY COOK DEVICE CAUSED OR CONTRIBUTED TO; OR IS LIKELY TO CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY IF A MALFUNCTION OCCURRED. BLANK FIELDS ON THIS FORM INDICATE THE INFORMATION IS UNKNOWN OR UNAVAILABLE.
ADDITIONAL INFORMATION: INVESTIGATION: H6 DEVICE CODE(S): APPROPRIATE TERM/CODE NOT AVAILABLE (3191) WAS SELECTED FOR THE ALLEGED DEVICE PERFORATION AND TILT. INVESTIGATION IS REOPENED DUE TO ADDITIONAL INFORMATION PROVIDED. THE REPORTED ALLEGATIONS HAVE BEEN FURTHER INVESTIGATED BASED ON THE INFORMATION PROVIDED TO DATE. VENA CAVA WALL PENETRATION/PERFORATION HAS BEEN REPORTED AND MAY BE EITHER SYMPTOMATIC OR ASYMPTOMATIC. POTENTIAL CAUSES MAY INCLUDE IMPROPER DEPLOYMENT; AND (OR) EXCESSIVE FORCE OR MANIPULATIONS NEAR AN IN-SITU FILTER (E.G., A SURGICAL OR ENDOVASCULAR PROCEDURE IN THE VICINITY OF A FILTER). POTENTIAL ADVERSE EVENTS THAT MAY OCCUR INCLUDE, BUT ARE NOT LIMITED TO, THE FOLLOWING: TRAUMA TO ADJACENT STRUCTURES, VASCULAR TRAUMA, VENA CAVA PERFORATION, VENA CAVA PENETRATION. FILTER TILT HAS BEEN REPORTED. POTENTIAL CAUSES MAY INCLUDE FILTER PLACEMENT IN IVCS WITH DIAMETERS LARGER THAN THOSE SPECIFIED IN THESE INSTRUCTIONS FOR USE; IMPROPER DEPLOYMENT; MANIPULATIONS NEAR AN IMPLANTED FILTER (E.G., A SURGICAL OR ENDOVASCULAR PROCEDURE IN THE VICINITY OF A FILTER); AND (OR) A FAILED RETRIEVAL ATTEMPT. EXCESSIVE FILTER TILT MAY CONTRIBUTE TO DIFFICULT OR FAILED RETRIEVAL; VENA CAVA WALL PENETRATION/PERFORATION; AND (OR) RESULT IN LOSS OF FILTER EFFICIENCY. POTENTIAL ADVERSE EVENTS THAT MAY OCCUR INCLUDE, BUT ARE NOT LIMITED TO, THE FOLLOWING: UNACCEPTABLE FILTER TILT. CATALOG NUMBER AND LOT NUMBER ARE UNKNOWN; HOWEVER, THERE IS NO EVIDENCE TO SUGGEST THAT THIS DEVICE WAS NOT MANUFACTURED ACCORDING TO SPECIFICATIONS AND NOTHING INDICATES THAT THE FILTER DID NOT PERFORM AS INTENDED, E.G. INTENDED FOR THE PREVENTION OF RECURRENT PULMONARY EMBOLISM (PE) VIA PLACEMENT IN THE VENA CAVA. THE FOLLOWING ALLEGATIONS HAVE BEEN INVESTIGATED: VENA CAVA PERFORATION AND TILT. NO EVIDENCE TO SUGGEST THAT THIS DEVICE WAS NOT MANUFACTURED ACCORDING TO SPECIFICATIONS AND NOTHING INDICATES THAT THE FILTER DID NOT PERFORM AS INTENDED, E.G. INTENDED FOR THE PREVENTION OF RECURRENT PULMONARY EMBOLISM (PE) VIA PLACEMENT IN THE VENA CAVA. COOK WILL REOPEN ITS INVESTIGATION IF FURTHER INFORMATION IS RECEIVED WARRANTING SUPPLEMENTATION IN ACCORDANCE WITH 21 C.F.R. 803.56. THIS REPORT INCLUDES INFORMATION KNOWN AT THIS TIME. A FOLLOW-UP MEDWATCH REPORT WILL BE SUBMITTED IF ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE. THIS REPORT IS REQUIRED BY THE FDA UNDER 21 CFR PART 803. THIS REPORT IS BASED ON UNCONFIRMED INFORMATION SUBMITTED BY OTHERS. NEITHER THE SUBMISSION OF THIS REPORT NOR ANY STATEMENT MADE IN IT IS INTENDED TO BE AN ADMISSION THAT ANY COOK DEVICE IS DEFECTIVE OR MALFUNCTIONED, THAT A DEATH OR SERIOUS INJURY OCCURRED, OR THAT ANY COOK DEVICE CAUSED OR CONTRIBUTED TO, OR IS LIKELY TO CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY IF A MALFUNCTION OCCURRED.
ADDITIONAL INFORMATION PROVIDED DETERMINED THAT THIS DEVICE WAS MANUFACTURED BY COOK INC. WITH THE SUBMISSION OF THIS FOLLOW UP REPORT, WILLIAM COOK EUROPE INFORMS THAT THIS COMPLAINT HAS BEEN TRANSFERRED FROM WILLIAM COOK EUROPE TO COOK INC.
PER REPORT FROM CT (COMPUTED TOMOGRAPHY): "POSITIVE FOR CAVAL PERFORATION. SUPERIOR EXTENT OF IVC FILTER MID BODY OF L1 AND INFERIOR EXTENT OF L2-L3 DISC SPACE. AT LEAST FOUR PRONGS HAVE PERFORATED THE IVC. MAXIMUM PERFORATION IS 9 MM SEEN ON SAGITTAL IMAGES. NO TILT OF CAVAL FILTER ON CORONAL IMAGES. A 6 MM ANTERIOR TO POSTERIOR TILT OF CAVAL FILTER ON SAGITTAL REFORMATTED IMAGES."
(B)(4). OCCUPATION: NON-HEALTHCARE PROFESSIONAL. PMA/510(K) K172557. IT HAS NOT BEEN POSSIBLE TO INVESTIGATE OR EVALUATE THIS ALLEGED EVENT BASED ON THE LIMITED INFORMATION PROVIDED TO DATE. COOK WILL REOPEN ITS INVESTIGATION IF FURTHER INFORMATION IS RECEIVING WARRANTING SUPPLEMENTATION IN ACCORDANCE WITH 21 C.F.R. 803.56.
'IT IS ALLEGED THAT "[PT] RECEIVED A COOK GUNTHER TULIP FILTER ON (B)(6) 2013". IT IS ALLEGED THAT [PT] WAS INJURED WITHOUT FURTHER EXPLANATION. HOSPITAL AND MEDICAL RECORDS HAVE BEEN REQUESTED BUT NOT YET PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 956422 | UNKNOWN | DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR | DTK | WILLIAM COOK EUROPE | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Life Threatening |