TRAM451N
Report
- Report Number
- 2124823-2007-00021
- Event Type
- Malfunction
- Date Received
- March 14, 2007
- Date of Event
- January 26, 2007
- Report Date
- March 14, 2007
- Manufacturer
- GE HEALTHCARE
- Product Code
- DRT
- PMA / PMN Number
- k011000
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- RISK MANAGER
Narratives
PART RETURNED TO GE FOR INVESTIGATION AND IT WAS DETERMINED THAT THE CARE UNIT IS UTILIZING WRONG SPO2 INTERCONNECT CABLE P/N 407252-002 WITH TRAM 451N MODULES. SPO2 INTERCONNECT CABLE P/N 407252-002 IS INCOMPATIBLE WITH THE TRAM 451N MODULES AND MAY RESULT IN INCORRECT SPO2 VALUES. THE P/N 407252-002 CABLE CONNECTOR IS KEYED AND THIS KEY DOES NOT MATCH THE KEYING ON THE 451N OXISMART TRAM MODULE. IN ORDER TO CONNECT THE CABLE TO THE MONITOR, THE KEY WOULD NEED TO BE DEFEATED (I.E. BROKEN). GE HEALTHCARE CONFIRMED THAT BOTH NEONATAL ICU AND HOSP HAVE BEEN PURGED OF INCOMPATIBLE SPO2 INTERCONNECT CABLE P/N 407252-002. CABLES HAVE BEEN REPLACED WITH COMPATIBLE 2006644-001 NELLCOR SPO2 INTERCONNECT CABLE. SPO2 TRAM MODULE AND CABLE COMPATIBILITY IS ADEQUATELY DESCRIBED IN LABELING VIA THE SOLAR 8000M PT MONITOR OPERATOR'S MANUAL, VERSION 4, 200701-107 CHAPTER 14, "SPO2": SECTION "INTRODUCTION", SUBSECTION, "MODULE AND PROBE COMPATIBILITY", "WARNING", "TRAM 451N AND TRAM 851N MODULES REQUIRE NELLCOR OXISMART XL CABLES AND PROBES. OLDER (NON-OXISMART XL) CABLES MUST NOT BE PLUGGED INTO THE SPO2 CONNECTOR ON THESE MODULES. USE OF NON-OXISMART XL CABLES MAY RESULT IN ERRONEOUS READINGS." AND "NOTE"; THE SPO2 CABLE SHOULD PLUG INTO THE MODULE'S SPO2 CONNECTOR EASILY AND SECURELY. DO NOT USE EXCESSIVE FORCE TO CONNECT THE CABLE. IF THE SPO2 CABLE DOES NOT EASILY FIT INTO THE SPO2 CONNECTOR ON THE MODULE, IT IS LIKELY THAT YOU DO NOT HAVE THE APPROPRIATE CABLE FOR THAT MODULE.
REPORT STATED "MARQUETTE MONITOR PULSE OXIMETER READING WAS BETWEEN 99 -100, WITH HEART RATE BETWEEN 190 - 192, BUT BABY CRYING, IRRITABLE, AND DUSKY IN COLOR. NURSE PLACED BABY ON ANOTHER BRAND OF PULSE OXIMETER MONITOR, SATURATION READING WAS BETWEEN 71 - 73. NURSE FELT THIS TO BE MORE REFLECTIVE OF BABY'S CLINICAL REPRESENTATION."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TRAM451N | PHYSIOLOGIC PATIENT MONITOR | DRT | GE HEALTHCARE | TRAM 451N | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 4 MO |