INTELLIS
Report
- Report Number
- 3004209178-2019-18996
- Event Type
- Malfunction
- Date Received
- October 4, 2019
- Date of Event
- October 1, 2019
- Report Date
- November 12, 2019
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- LGW
- UDI-DI
- 00643169781702
- PMA / PMN Number
- P840001
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- 003
Narratives
CONCOMITANT MEDICAL PRODUCTS: PRODUCT ID (B)(4), SERIAL# (B)(6), IMPLANTED: (B)(6) 2019, PRODUCT TYPE: LEAD, PRODUCT ID (B)(4), SERIAL# (B)(6), IMPLANTED: (B)(6) 2019, PRODUCT TYPE: LEAD. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
ADDITIONAL INFORMATION RECEIVED FROM THE MANUFACTURER REPRESENTATIVE REPORTED THAT THE CAUSE WAS NOT DETERMINED. IT WAS UNKNOWN IF THE PATIENT HAD OBTAINED PAIN RELIEF AS THERE HAD BEEN NO CONTACT SINCE THE LAST MEETING.
CONCOMITANT MEDICAL PRODUCTS: PRODUCT ID: 977A260, SERIAL#: (B)(4), IMPLANTED: (B)(6) 2019, PRODUCT TYPE: LEAD. PRODUCT ID: 977A260, SERIAL#: (B)(4), IMPLANTED: (B)(6) 2019, PRODUCT TYPE: LEAD. OTHER RELEVANT DEVICE(S) ARE: PRODUCT ID: 977A260, SERIAL/LOT#: (B)(4), UBD: 08-MAY-2023, UDI#: (B)(4); PRODUCT ID: 977A260, SERIAL/LOT #: (B)(4), UBD: 12-APR-2023, UDI#: (B)(4). IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
INFORMATION WAS RECEIVED FROM A PATIENT AND MANUFACTURING REPRESENTATIVE (REP) REGARDING THE PATIENT'S IMPLANTABLE NEUROSTIMULATOR (INS). INFORMATION WAS REPORTED THAT THE PATIENT HAS NO PAIN RELIEF. X-RAYS WERE TAKEN AND LEADS HAVE MIGRATED FROM THE TOP OF T8 TO THE BOTTOM OF T9 AND T9/10 INTERSPACE. THE PATIENT WAS REPROGRAMMED TO TARGET T9/10 INTERSPACE. IT IS UNKNOWN AT THE TIME OF THE REPORT IF THE ISSUE HAS BEEN RESOLVED. NO FURTHER COMPLICATIONS WERE REPORTED. NO ADDITIONAL PATIENT SYMPTOMS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 951207 | INTELLIS | STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF | LGW | MEDTRONIC PUERTO RICO OPERATIONS CO. | 97715 | 00643169781702 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 79 YR |