FDA Adverse Event Malfunction Summary report: N

COBAS B 101 HBA1C TEST

MDR report key: 9152603 · Received October 3, 2019

Report

Report Number
1823260-2019-03562
Event Type
Malfunction
Date Received
October 3, 2019
Date of Event
September 5, 2019
Report Date
November 20, 2019
Manufacturer
ROCHE DIAGNOSTICS
Product Code
LCP
PMA / PMN Number
K163633
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

RETENTION COBAS B101 HBA1C TEST DISC LOT 929042-01 AND LOT 930041-01 WERE INVESTIGATED. THE MANUFACTURER BATCH RECORD FOR COBAS B101 HBA1C TEST DISC LOT 929042-01 AND LOT 930041-01 SHOWED NO ABNORMALITIES. THE MANUFACTURER FINAL PRODUCT INSPECTION REPORT FOR COBAS B101 HBA1C TEST DISC LOT 929042-01 AND LOT 930041-01 SHOWED NO ABNORMALITIES AND ALL FEATURES PASSED. RETENTION DISCS COBAS B101 HBA1C DISC LOT 929042-01 AND LOT 930041-01 WERE MEASURED WITH BLOOD COMPARED TO REFERENCE METHOD: 4 REPLICATES 6 REPLICATES AT 2 LEVELS (WHOLE BLOOD LOW AND MIDDLE): 4 WHOLE BLOOD "LOW" SAMPLES WERE TESTED WITH DISCS RETENTION MATERIAL HBA1C LOT 929042-01 WITH RETENTION PHC COBAS B101 INSTRUMENT. RESULTS DID NOT SHOW THE DIFFERENCE WHICH THE CUSTOMER IS CLAIMING. 4 WHOLE BLOOD "MIDDLE" SAMPLES WERE TESTED WITH DISCS RETENTION MATERIAL HBA1C LOT 929042-01 WITH RETENTION PHC COBAS B101 INSTRUMENT. RESULTS DID NOT SHOW THE DIFFERENCE WHICH THE CUSTOMER IS CLAIMING. 6 WHOLE BLOOD "LOW" SAMPLES WERE TESTED WITH DISCS RETENTION MATERIAL HBA1C LOT 930041-01 WITH RETENTION PHC COBAS B101 INSTRUMENT. RESULTS DID NOT SHOW THE DIFFERENCE WHICH THE CUSTOMER IS CLAIMING. 6 WHOLE BLOOD "MIDDLE" SAMPLES WERE TESTED WITH DISCS RETENTION MATERIAL HBA1C LOT 930041-01 WITH RETENTION PHC COBAS B101 INSTRUMENT. RESULTS DID NOT SHOW THE DIFFERENCE WHICH THE CUSTOMER IS CLAIMING. THERE WERE NO DISCREPANT HBA1C RESULTS MEASURED WITH RETENTION MATERIAL HBA1C DISC LOT 929042-01 AND LOT 930041-01. DISCREPANT HBA1C RESULTS COULD NOT BE CONFIRMED. THE PROVIDED CUSTOMER DATA SHOWED DATA DEVIATION BETWEEN 1ST AND 2ND MEASUREMENT WAS 0.6-1.0% NGSP ON HBA1C LOT 929042-01. - RETENTION MATERIAL DATA SHOWED DATA DEVIATION BETWEEN 1ST AND 2ND MEASUREMENT WAS 0-0.1% NGSP ON HBA1C LOT 929042-01. - RETENTION MATERIAL TESTING: DATA DEVIATION BETWEEN HBA1C TEST DISC LOT 929042-01 AND LOT 930041-01 IS 0.1% NGSP. THE PROVIDED CUSTOMER'S DATA DEVIATION WAS 1.0%NGSP. - THE INVESTIGATION SHOWED THAT THE MEASUREMENT DATA OBTAINED WITH RETENTION MATERIAL HAS NO DEVIATION LIKE THE CUSTOMER COMPLAINED ABOUT. UPON REVIEW OF THE INSTRUMENT'S DATA LOGS, A WARNING MESSAGE OCCURRED AT THE SAME TIME THE COMPLAINED MEASUREMENTS WERE PERFORMED FOR THE FIRST SAMPLE. THE MESSAGE INDICATES ABNORMALLY HIGH HEMOGLOBIN. THIS MESSAGE OCCURS IF THE B101 INSTRUMENT DETECTS THAT THE HEMOGLOBIN CONCENTRATION IN THE TESTED BLOOD SAMPLE IS OVER THE MEASUREABLE RANGE (20G/DL). THE MESSAGE CAN ALSO OCCUR IF THE BLOOD SAMPLE ENTERED INTO THE DISC THROUGH THE INADEQUATE PORTION (HOLE). THE MESSAGE CAN ALSO OCCUR IF THE SAMPLE VOLUME IS TOO MUCH WHEN THE SAMPLE IS OVERFLOWED FROM THE CAPILLARY CHANNEL DUE TO THE IMPACT SUCH AS A DROP, THE HEMOGLOBIN CONCENTRATION IS OVER THE MEASURABLE RANGE. THE SAME WARNING MESSAGE OCCURRED MANY TIMES BETWEEN (B)(6) 2019 AND (B)(6) 2019.

Additional Manufacturer Narrative · 1

THE REPORTER'S PRODUCT WAS REQUESTED FOR INVESTIGATION. THIS EVENT OCCURRED IN (B)(6).

Description of Event or Problem · 1

THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR THREE PATIENT SAMPLES TESTED WITH THE COBAS B 101 HBA1C TEST ON COBAS B 101 INSTRUMENT SERIAL NUMBER (B)(4). IT WAS ASKED, BUT IT IS NOT KNOWN IF ANY INCORRECT RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY. THE SAMPLES WERE REPEATED AND THE REPEAT MEASUREMENTS MATCHED THE CLINICAL CONDITION AND PREVIOUS RESULTS OF THE PATIENTS. THE REPORTER ALSO MENTIONED THAT THEY RECENTLY RECEIVE FREQUENT ERRORS ON THE COBAS B 101 INSTRUMENT. THE FIRST SAMPLE INITIALLY RESULTED WITH AN HBA1C VALUE OF 8.4 %, WHICH REPEATED AS 7.4 %. BOTH MEASUREMENTS WERE PERFORMED USING HBA1C TEST DISC LOT 929042-01. THE SECOND SAMPLE INITIALLY RESULTED WITH AN HBA1C VALUE OF 6.6 %, WHICH REPEATED AS 6.0 %. BOTH MEASUREMENTS WERE PERFORMED ON (B)(6) 2019 USING HBA1C TEST DISC LOT 929042-01. THE THIRD SAMPLE INITIALLY RESULTED WITH AN HBA1C VALUE OF 6.5 % WHEN MEASURED USING HBA1C TEST DISC LOT 929042-01 ON (B)(6) 2019. THE SAMPLE WAS REPEATED ON (B)(6) 2019 USING HBA1C TEST DISC LOT 930041-01, RESULTING WITH A VALUE OF 5.5 %.