FDA Adverse Event Malfunction Summary report: N

SYR 10 ML FIL SALINE SHORTLENGTH PLUNGER ROD

MDR report key: 9152044 · Received October 3, 2019

Report

Report Number
1911916-2019-01043
Event Type
Malfunction
Date Received
October 3, 2019
Date of Event
September 13, 2019
Report Date
October 11, 2019
Manufacturer
BECTON DICKINSON AND COMPANY
Product Code
NGT
UDI-DI
00382903064991
PMA / PMN Number
K161552
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: ONE SAMPLE WAS RECEIVED. IT HAS THE PLUNGER ROD-RUBBER STOPPER, NO TIP CAP, AND NO SALINE SOLUTION; THE PLUNGER ROD-RUBBER STOPPER IS ALL THE WAY DOWN. THE PLUNGER ROD IS NOT ATTACHED TO THE RUBBER STOPPER. NO MANUFACTURING ISSUES WERE EXPERIENCED DURING THE PRODUCTION OF THESE PRODUCTS. THE SYRINGE IS NOT DESIGNED TO DRAWING BLOOD. POSSIBLE ROOT CAUSE, OFF LABEL USE. THE SYRINGE IS DESIGNED TO FLUSH THE IV LINES BY PUSHING THE PLUNGER ROD DOWN, NOT TO DRAW BLOOD LAB SAMPLES BY PULLING THE PLUNGER ROD UP. FURTHERMORE, A DEVICE HISTORY RECORD REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION OF THE PROVIDED LOT NUMBER THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. H3 OTHER TEXT : SEE H.10.

Description of Event or Problem · 0

IT HAS BEEN REPORTED THAT ONE SYR 10 ML FIL SALINE SHORTLENGTH PLUNGER ROD HAS BEEN FOUND EXPERIENCING STOPPER SEPARATION FROM THE PLUNGER DURING USE. THE FOLLOWING HAS BEEN PROVIDED BY THE INITIAL REPORTER: IT WAS REPORTED THAT THE "BLACK RUBBER ON NS SYRINGE BECAME DISCONNECTED FROM PLUNGER WHILE TRYING TO WITHDRAW MY WASTE DURING LAB DRAW." CUSTOMER TEXT: CONCERN: BLACK RUBBER ON NS SYRINGE BECAME DISCONNECTED FROM PLUNGER WHILE TRYING TO WITHDRAW MY WASTE DURING LAB DRAW. I UNHOOKED THE SYRINGE AND HEPLOCKED THE IV AND GRABBED NEW SUPPLIES TO START OVER--AND HAD NO PROBLEMS.

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT HAS BEEN REPORTED THAT ONE SYR 10 ML FIL SALINE SHORTLENGTH PLUNGER ROD HAS BEEN FOUND EXPERIENCING STOPPER SEPARATION FROM THE PLUNGER DURING USE. THE FOLLOWING HAS BEEN PROVIDED BY THE INITIAL REPORTER: IT WAS REPORTED THAT THE "BLACK RUBBER ON NS SYRINGE BECAME DISCONNECTED FROM PLUNGER WHILE TRYING TO WITHDRAW MY WASTE DURING LAB DRAW." CUSTOMER TEXT: CONCERN: BLACK RUBBER ON NS SYRINGE BECAME DISCONNECTED FROM PLUNGER WHILE TRYING TO WITHDRAW MY WASTE DURING LAB DRAW. I UNHOOKED THE SYRINGE AND HEPLOCKED THE IV AND GRABBED NEW SUPPLIES TO START OVER--AND HAD NO PROBLEMS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
948223 SYR 10 ML FIL SALINE SHORTLENGTH PLUNGER ROD INTRAVASCULAR CATHETER NGT BECTON DICKINSON AND COMPANY 9018603 00382903064991

Patients

Seq Age Sex Outcome Treatment
1 Other