ELECSYS ¿-CROSSLAPS/SERUM IMMUNOASSAY
Report
- Report Number
- 1823260-2019-03556
- Event Type
- Malfunction
- Date Received
- October 3, 2019
- Date of Event
- May 31, 2019
- Report Date
- February 4, 2020
- Manufacturer
- ROCHE DIAGNOSTICS
- Product Code
- JMM
- PMA / PMN Number
- K993706
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE QC DATA PROVIDED FROM THE E801 MODULE WAS ACCEPTABLE. A SAMPLE WAS RETURNED FOR INVESTIGATION. THE INVESTIGATION CONFIRMED THE CUSTOMER¿S RESULTS. A HETEROPHILIC INTERFERENCE WAS FOUND IN THE SAMPLE. THE INTERFERENCE IS DESCRIBED IN THE PRODUCT LABELING: IN RARE CASES, INTERFERENCE DUE TO EXTREMELY HIGH TITERS OF ANTIBODIES TO ANALYTE-SPECIFIC ANTIBODIES, STREPTAVIDIN OR RUTHENIUM CAN OCCUR. THESE EFFECTS ARE MINIMIZED BY SUITABLE TEST DESIGN.
THIS EVENT OCCURRED IN (B)(6). THE CUSTOMER PERFORMED A NEUTRALIZATION PROCEDURE ON THE PATIENT SAMPLE IN AN ATTEMPT TO ELIMINATE THE INTERFERENCE WITH NO SUCCESS.
THE INITIAL REPORTER COMPLAINED OF DISCREPANT HIGH RESULTS FOR 1 PATIENT TESTED FOR ELECSYS B-CROSSLAPS/SERUM (B-CROSSLAPS) ON A COBAS E 801 MODULE COMPARED TO A COBAS E 411 IMMUNOASSAY ANALYZER. THE INITIAL RESULT FROM THE E801 MODULE WAS > 6000 PG/ML. SINCE THIS RESULT DID NOT CORRESPOND TO THE PATIENT'S CLINICAL PICTURE, A NEW SAMPLE WAS REQUESTED. ON (B)(6) 2019 A NEW SAMPLE WAS OBTAINED AND THE RESULT ON THE E801 MODULE WAS 5598 PG/ML. THIS SAMPLE WAS REPEATED ON AN E411 ANALYZER IN A DIFFERENT LABORATORY AND THE RESULT WAS 389.4 PG/ML. THIS RESULT WAS BELIEVED TO BE CORRECT. THE CUSTOMER SUSPECTS AN INTERFERENCE OF THE PATIENT SAMPLE AND THE E801 MODULE METHODOLOGY. THE E801 MODULE SERIAL NUMBER IS (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 948320 | ELECSYS ¿-CROSSLAPS/SERUM IMMUNOASSAY | COLUMN CHROMATOGRAPHY & COLOR DEVELOPMENT, HYDROXYPROLINE | JMM | ROCHE DIAGNOSTICS | NA | 349942 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 77 YR |