G7 PPS LTD ACET SHELL 44A
Report
- Report Number
- 0001825034-2019-04419
- Event Type
- Malfunction
- Date Received
- October 2, 2019
- Date of Event
- September 25, 2019
- Report Date
- January 29, 2020
- Manufacturer
- ZIMMER BIOMET, INC.
- Product Code
- PBI
- PMA / PMN Number
- K121874
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PL
- Reporter Occupation
- PHYSICIAN
Narratives
UDI#: (B)(4). THE EVENT WAS CONFIRMED WITH PRODUCT RECEIVED. VISUAL INSPECTION FOUND LIGHT SCRATCHES ON THE INNER RADIUS AND RIM OF THE SHELL. NO DAMAGE WAS OBSERVED TO THE LOCKING GROOVE OR ORIENTATION FEATURE. THE APICAL PLUG WAS ALSO RETURNED WITH THE COMPLAINT. DHR WAS REVIEWED AND NO DISCREPANCIES RELEVANT TO THE REPORTED EVENT WERE FOUND. ROOT CAUSE WAS UNABLE TO BE DETERMINED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
(B)(4). CONCOMITANT MEDICAL PRODUCT: CATALOG# 010000757, G7 10 DEG ARCOMXL LINER 28MM A, LOT# 3882068. CATALOG# 010000913, G7 HI-WALL E1 LINER 28MM A, LOT# 6167725. CATALOG# 010000874, G7 10 DEG E1 LINER 28MM A, LOT# 6569774. CATALOG# 110003625, BIOLOX DELTA CER LNR 28MM B, LOT# 6022564. CATALOG# 650-0830, DELTA CER FM HD 028/-3.5 12/14, LOT# 2019020164. CATALOG# 010000660, G7 PPS LTD ACET SHELL 46B, LOT# 6360904. CATALOG# 650-0951, MICRO TPRLC STD PC 6MM (12/14), LOT# 6594650. REPORT SOURCE : (B)(6). CUSTOMER HAS INDICATED THAT THE PRODUCT IS IN PROCESS OF BEING RETURNED TO ZIMMER BIOMET FOR INVESTIGATION. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED. MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORTS: 0001825034-2019-04417, 0001825034-2019-04418, 0001825034-2019-04420.
IT WAS REPORTED THAT THE LINER WOULD NOT LOCK IN THE ACETABULAR SHELL. SURGEON USED ANOTHER, LARGER SHELL AND LINER TO COMPLETE THE SURGERY. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 940449 | G7 PPS LTD ACET SHELL 44A | PROSTHESIS, HIP | PBI | ZIMMER BIOMET, INC. | N/A | 6129679 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |