OMNIPOD DASH¿ INSULIN MANAGEMENT SYSTEM
Report
- Report Number
- 3004464228-2019-09464
- Event Type
- Malfunction
- Date Received
- October 1, 2019
- Date of Event
- September 19, 2019
- Report Date
- September 19, 2019
- Manufacturer
- INSULET CORPORATION
- Product Code
- LZG
- UDI-DI
- 20385082000020
- PMA / PMN Number
- K180045
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER
Narratives
THE RECEIVED DEVICE HAD THE CANNULA ASSEMBLY DEPLOYED. INSPECTION OF THE FLUID PATH FOUND THE EXPOSED PORTION OF THE SOFT CANNULA BENT. ALTHOUGH THIS DAMAGE WAS OBSERVED, IT CANNOT BE DETERMINED WHEN OR HOW THIS DAMAGE OCCURRED. NO OTHER DEFECTS OR DEFICIENCIES WERE FOUND THAT WOULD RESULT IN A FAILURE OF THE DEVICE TO DELIVER INSULIN CORRECTION TO : EXPIRATION DATE CHANGED FROM 1021672 TO 1081672. UNIQUE IDENTIFIER (UDI) # CHANGED FROM (B)(4) TO ((B)(4).
THE DEVICE WAS NOT RETURNED FOR EVALUATION. WE ARE UNABLE TO CONFIRM THE BENT CANNULA OR TO DETERMINE IF IT COULD HAVE CONTRIBUTED TO THE REPORTED HYPERGLYCEMIA. LOT RELEASE RECORDS WERE REVIEWED AND THE PRODUCT LOT MET ALL ACCEPTANCE CRITERIA. OMNIPOD DASH INSULIN MANAGEMENT SYSTEM ¿ USER GUIDE. MODEL: 18320. 18296-ENG-AW REV B 06/18. BLOOD GLUCOSE READINGS. CHAPTER 4 / PAGE 56. WARNINGS: BLOOD GLUCOSE READINGS BELOW 70 MG/DL MAY INDICATE HYPOGLYCEMIA (LOW BLOOD GLUCOSE). BLOOD GLUCOSE READINGS ABOVE 250 MG/DL MAY INDICATE HYPERGLYCEMIA (HIGH BLOOD GLUCOSE). FOLLOW YOUR HEALTHCARE PROVIDER'S SUGGESTIONS FOR TREATMENT.
IT WAS REPORTED THAT PATIENT'S BG (BLOOD GLUCOSE) REACHED 531 MG/DL WHILE WEARING THE POD BETWEEN 24 AND 36 HOURS. THE PATIENT'S CANNULA WAS FOUND BENT WHEN REMOVED FROM THE INFUSION SITE (ARM). FOR TREATMENT, GAVE CORRECTIONS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 933951 | OMNIPOD DASH¿ INSULIN MANAGEMENT SYSTEM | PUMP, INFUSION, INSULIN | LZG | INSULET CORPORATION | 18320 | PD1C11161851 | 20385082000020 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 15 YR |