FDA Adverse Event Malfunction Summary report: N

INRATIO

MDR report key: 913213 · Received September 13, 2007

Report

Report Number
2954730-2007-00433
Event Type
Malfunction
Date Received
September 13, 2007
Date of Event
August 27, 2007
Report Date
September 11, 2007
Manufacturer
HEMOSENSE, INC.
Product Code
JPA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

DISCREPANT RESULTS (ACCURACY) COMPARISON OF INRATIO TEST WITH LAB RESULTS PROVIDED BY END-USER AT TIME COMPLAINT WAS FILED: PER INTERNAL PROCEDURE, THE MEAN OF THE INRATIO METER AND COMPARATIVE SYSTEM INR WERE CALCULATED. FOR ALL 3 DATA SETS, BOTH INRATIO AND LAB VALUES ARE WITHIN THE CONFIDENCE LIMITS FOR INR TESTING. THE RESULTS ARE NOT CONSIDERED DISCREPANT WITHIN THE CONTEXT OF THE DOCUMENTED VARIABILITY FOR INR TESTING. THEREFORE, FURTHER TESTING IS NOT REQUIRED AT THIS TIME. INRATIO PRECISION DATA PROVIDED BY END-USER LOT 070274: FIRST TEST INR = 1.2; SECOND TEST INR = 1.3; MEAN = 1.25; SD = 0.07; %CV = 5.66%. THE %CV IS LESS THAN OR EQUAL TO 20%. PER INTERNAL PROCEDURE, THE PRECISION PASSES THE CRITERIA FOR PRECISION. THE TEST IS CONSIDERED PRECISE AND NO FURTHER TESTING IS REQUIRED AT THIS TIME.

Description of Event or Problem · 1

CALLER ALLEGES INACCURACY WITH INRATIO. RESULTS AS FOLLOWS: CALLER ALLEGES IMPRECISION WITH INRATIO. RESULTS AS FOLLOWS: FIRST TEST INR = 1.2; SECOND TEST INR = 1.3.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INRATIO PROTHROMBIN TIME TEST JPA HEMOSENSE, INC. 070274

Patients

Seq Age Sex Outcome Treatment
1 YR