MAESTRO RADIAL STEM 4.5X40MM
Report
- Report Number
- 0001825034-2019-04291
- Event Type
- Injury
- Date Received
- September 27, 2019
- Report Date
- November 25, 2019
- Manufacturer
- ZIMMER BIOMET, INC.
- Product Code
- JWJ
- PMA / PMN Number
- K042032
- Removal / Correction Number
- N/A
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). REPORTED EVENT WAS CONSIDERED CONFIRMED DUE TO THE PROVED X-RAYS THAT SHOWED LOOSENING OF THE RADIAL COMPONENT. DEVICE HISTORY RECORD WAS REVIEWED AND NO DISCREPANCIES RELEVANT TO THE REPORTED EVENT WERE FOUND. ROOT CAUSE WAS CONSIDERED TO BE USER ERROR AS IT WAS KNOWN THAT THE PATIENT WAS A NURSE THAT LIFTED PATIENTS. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS. MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORTS: 0001825034-2019-05314
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
(B)(4). (UDI): N/A. MEDICAL PRODUCT: CATALOG #: 180151, MAESTRO RAD W/BRG 7X15 7.5 RT, LOT # 922440; CATALOG #: 180351, VARIABLE LOCK SCREW 4.75X20MM, LOT # 748570; CATALOG #: 180363, MAESTRO TC CARPALHD 7X15MM STD, LOT # 527190; CATALOG #: 180352, VARIABLE LOCK SCREW 4.75X25MM, LOT # 551330; CATALOG #: 180363, MAESTRO TC CARPALHD 7X15MM STD, LOT # 527190; CATALOG #: 180321, MAESTRO TC CAPITATESTEM 6X15MM, LOT # 238690. THE CUSTOMER HAS NOT INDICATED WHETHER THE PRODUCT WILL BE RETURNED TO ZIMMER BIOMET FOR INVESTIGATION OR NOT. ONCE THIS INFORMATION IS OBTAINED A FOLLOW-UP MDR WILL BE SUBMITTED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS. UNKNOWN IF PRODUCT WILL BE RETURNED.
IT WAS REPORTED THAT THE PATIENT IS SCHEDULED TO BE GETTING A REVISION DUE TO THE RADIAL STEM BEING LOOSE IN THE CANAL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 921556 | MAESTRO RADIAL STEM 4.5X40MM | PROSTHESIS SHOULDER | JWJ | ZIMMER BIOMET, INC. | N/A | 181430 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| R |