FDA Adverse Event Other Summary report: N

INTROCAN SAFETY

MDR report key: 911747 · Received September 12, 2007

Report

Report Number
2523676-2007-00055
Event Type
Other
Date Received
September 12, 2007
Date of Event
August 15, 2007
Report Date
August 31, 2007
Manufacturer
B. BRAUN MEDICAL, INC.
Product Code
FOZ
PMA / PMN Number
K982805
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE ACTUAL INTROCAN SAFETY NEEDLE WITH CLIP INVOLVED IN THE INCIDENT WAS RETURNED TO THE MFR TO BE EVALUATED. THE CLIP WAS LOCATED AT THE TIP OF THE NEEDLE. HOWEVER, THE CLIP DID NOT COVER THE TIP OF THE NEEDLE. THE LONG ARM OF THE CLIP WAS PUSHED OFF THE SIDE OF THE NEEDLE. THE CATHETER WAS NOT RETURNED FOR EVAL. ON 09/11/07 - THE FACILITY WAS CONTACTED TO INFORM THEM THAT THE RETURNED SAMPLE DID NOT HAVE THE CLIP COVERING THE TIP OF THE NEEDLE AS INITIALLY REPORTED. THE REPORTER INDICATED THAT THE NURSE DID NOT RECEIVE A NEEDLESTICK. THE NURSE DISTINCTLY FELT THE SIDE OF THE CLIP PUNCTURE HER SKIN. ALL SAFETY CLIP DIMENSIONS THAT WERE ABLE TO BE MEASURED ON THE RETURNED SAMPLE WERE CHECKED AND FOUND TO BE WITHIN THE DESIGN SPECIFICATIONS. THE SAFETY CLIP IS DESIGNED TO COVER THE TIP OF THE NEEDLE, HOWEVER, THIS PORTION OF THE NEEDLE IS MADE OF RIGID METAL AND SHOULD STILL NOT BE HANDLE AFTER THE CLIP IS ACTIVATED. CDC GUIDELINES AND/OR FACILITY PROTOCOLS SHOULD ALWAYS BE FOLLOWED. SHARPS SHOULD BE DISPOSED OF IMMEDIATELY INTO AN APPROPRIATE SHARPS CONTAINER. THE SAMPLE AND ALL AVAILABLE INFO HAS BEEN PROVIDED TO THE ACTUAL MFR, B. BRAUN MEDICAL INDUSTRIES SDN BHD, IN MALAYSIA.

Description of Event or Problem · 1

AS REPORTED BY THE USER FACILITY: NURSE USING INTROCAN. INSERTION AND REMOVAL OF THE STYLET WITHOUT INCIDENT, SAFETY CLIP ENGAGED. REPORTED SAFETY CLIP PUNCTURED SKIN. ADDITIONAL INFO PROVIDED BY THE FACILITY INDICATED THE NURSE IS FINE AND ALL PROTOCOL BLOOD WORK TESTING IS NEGATIVE. THE LOT NUMBER REMAINS UNKNOWN. NO FURTHER INFO IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INTROCAN SAFETY I.V. SAFETY CATHETER FOZ B. BRAUN MEDICAL, INC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1 YR Other