FDA Adverse Event Malfunction Summary report: N

BD SYRINGE 50ML PERFUSION

MDR report key: 9112305 · Received September 24, 2019

Report

Report Number
3003152976-2019-00662
Event Type
Malfunction
Date Received
September 24, 2019
Date of Event
September 8, 2019
Report Date
October 10, 2019
Manufacturer
BECTON DICKINSON, S.A.
Product Code
FMF
PMA / PMN Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: NO PHOTOS OR PHYSICAL SAMPLES THAT DISPLAY THE REPORTED CONDITION WERE AVAILABLE FOR INVESTIGATION. A DEVICE HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT 1902205, NO DEVIATIONS OR NON-CONFORMANCES WERE IDENTIFIED DURING THE MANUFACTURING PROCESS THAT COULD HAVE CONTRIBUTED TO THIS ISSUE. TEN RETAINED SAMPLES OF LOT 1902205 WERE USED TO CONDUCT A LEAKAGE TEST. THE PRODUCT WAS VISUALLY INSPECTED, NO DEFECTS OR DAMAGE WAS NOTED, AND NO LEAK WAS IDENTIFIED. BASED ON THE AVAILABLE INFORMATION WE ARE NOT ABLE TO DETERMINE A ROOT CAUSE AT THIS TIME.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHEN FILLING DIFFUSER PUMP WITH SYRINGE, MEDICATION LEAKED WITH A BD SYRINGE 50ML PERFUSION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM FRENCH TO ENGLISH: THIS SYRINGE WAS USED TO FILL A CHEMOTHERAPY DIFFUSER PUMP (WHEREAS WE USUALLY USE ANOTHER TYPE OF SYRINGE, 60 ML CAPACITY). IN THIS DIFFUSER FILLING SITUATION, WE PLACE THE PISTON AT THE BOTTOM AGAINST THE BENCH AND PRESS UP AND DOWN TO FILL THE DIFFUSER. THIS CAUSED THE PRODUCT TO FLOW BACK THROUGH THE PISTON.

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT WHEN FILLING DIFFUSER PUMP WITH SYRINGE, MEDICATION LEAKED WITH A BD SYRINGE 50 ML PERFUSION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM FRENCH TO ENGLISH: THIS SYRINGE WAS USED TO FILL A CHEMOTHERAPY DIFFUSER PUMP (WHEREAS WE USUALLY USE ANOTHER TYPE OF SYRINGE, 60 ML CAPACITY). IN THIS DIFFUSER FILLING SITUATION, WE PLACE THE PISTON AT THE BOTTOM AGAINST THE BENCH AND PRESS UP AND DOWN TO FILL THE DIFFUSER. THIS CAUSED THE PRODUCT TO FLOW BACK THROUGH THE PISTON.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
907759 BD SYRINGE 50ML PERFUSION PISTON SYRINGE FMF BECTON DICKINSON, S.A. 1902205

Patients

Seq Age Sex Outcome Treatment
1 Other