EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE
Report
- Report Number
- 8010047-2019-03352
- Event Type
- Malfunction
- Date Received
- September 24, 2019
- Date of Event
- August 6, 2019
- Report Date
- October 17, 2019
- Manufacturer
- OLYMPUS MEDICAL SYSTEMS CORP.
- Product Code
- FDS
- UDI-DI
- 04953170305290
- PMA / PMN Number
- K131780
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
THE SUBJECT DEVICE HAS NOT BEEN RETURNED TO OMSC BUT WAS RETURNED TO OLYMPUS (B)(4). (B)(4) SENT THE DEVICE TO A THIRD PARTY LABORATORY FOR MICROBIOLOGICAL TESTING. AS A RESULT OF THE TESTING, NO MICROBE WAS DETECTED FROM THE SAMPLE COLLECTED FROM ALL CHANNELS OF THE SUBJECT DEVICE. THE TESTING RESULT CLEARED THE (B)(6) GUIDELINE. THE EXACT CAUSE OF THE REPORTED EVENT COULD NOT BE CONCLUSIVELY DETERMINED AT THIS TIME. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, THIS REPORT WILL BE SUPPLEMENTED.
THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE ADDITIONAL INFORMATION. OMSC REVIEWED THE MANUFACTURING HISTORY (DHR) OF THE SUBJECT DEVICE AND CONFIRMED NO IRREGULARITY. THE EXACT CAUSE OF THE REPORTED EVENT COULD NOT BE CONCLUSIVELY DETERMINED. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, THIS REPORT WILL BE SUPPLEMENTED.
OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) WAS INFORMED THAT AS A RESULT OF MICROBIOLOGICAL TESTING BY THE USER FACILITY, THE FOLLOWING MICROBES WERE DETECTED FROM THE SAMPLE COLLECTED FROM THE SUBJECT DEVICE. INSTRUMENT CHANNEL: PSEUDOMONAS AERUGINOSA (2 CFU) SUCTION CHANNEL: PSEUDOMONAS AERUGINOSA (5 CFU) THE DEVICE HAD BEEN REPROCESSED WITH A NON-OLYMPUS AUTOMATED ENDOSCOPE REPROCESSOR, SOLUSCOPE S4, USING PERACETIC ACID. THERE WAS NO REPORT OF INFECTION ASSOCIATED WITH THIS REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 902763 | EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE | GASTROINTESTINAL VIDEOSCOPE | FDS | OLYMPUS MEDICAL SYSTEMS CORP. | GIF-H190 | 04953170305290 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |