FDA Adverse Event Malfunction Summary report: N

EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE

MDR report key: 9108090 · Received September 24, 2019

Report

Report Number
8010047-2019-03352
Event Type
Malfunction
Date Received
September 24, 2019
Date of Event
August 6, 2019
Report Date
October 17, 2019
Manufacturer
OLYMPUS MEDICAL SYSTEMS CORP.
Product Code
FDS
UDI-DI
04953170305290
PMA / PMN Number
K131780
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THE SUBJECT DEVICE HAS NOT BEEN RETURNED TO OMSC BUT WAS RETURNED TO OLYMPUS (B)(4). (B)(4) SENT THE DEVICE TO A THIRD PARTY LABORATORY FOR MICROBIOLOGICAL TESTING. AS A RESULT OF THE TESTING, NO MICROBE WAS DETECTED FROM THE SAMPLE COLLECTED FROM ALL CHANNELS OF THE SUBJECT DEVICE. THE TESTING RESULT CLEARED THE (B)(6) GUIDELINE. THE EXACT CAUSE OF THE REPORTED EVENT COULD NOT BE CONCLUSIVELY DETERMINED AT THIS TIME. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, THIS REPORT WILL BE SUPPLEMENTED.

Additional Manufacturer Narrative · 1

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE ADDITIONAL INFORMATION. OMSC REVIEWED THE MANUFACTURING HISTORY (DHR) OF THE SUBJECT DEVICE AND CONFIRMED NO IRREGULARITY. THE EXACT CAUSE OF THE REPORTED EVENT COULD NOT BE CONCLUSIVELY DETERMINED. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, THIS REPORT WILL BE SUPPLEMENTED.

Description of Event or Problem · 1

OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) WAS INFORMED THAT AS A RESULT OF MICROBIOLOGICAL TESTING BY THE USER FACILITY, THE FOLLOWING MICROBES WERE DETECTED FROM THE SAMPLE COLLECTED FROM THE SUBJECT DEVICE. INSTRUMENT CHANNEL: PSEUDOMONAS AERUGINOSA (2 CFU) SUCTION CHANNEL: PSEUDOMONAS AERUGINOSA (5 CFU) THE DEVICE HAD BEEN REPROCESSED WITH A NON-OLYMPUS AUTOMATED ENDOSCOPE REPROCESSOR, SOLUSCOPE S4, USING PERACETIC ACID. THERE WAS NO REPORT OF INFECTION ASSOCIATED WITH THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
902763 EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE GASTROINTESTINAL VIDEOSCOPE FDS OLYMPUS MEDICAL SYSTEMS CORP. GIF-H190 04953170305290

Patients

Seq Age Sex Outcome Treatment
1