FDA Adverse Event Malfunction Summary report: N

ORBERA365 INTRAGASTRIC BALLOON SYSTEM

MDR report key: 9105789 · Received September 23, 2019

Report

Report Number
3006722112-2019-00096
Event Type
Malfunction
Date Received
September 23, 2019
Date of Event
December 28, 2018
Report Date
February 11, 2019
Manufacturer
APOLLO ENDOSURGERY, INC.
Product Code
LTI
PMA / PMN Number
P140008
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION REGARDING THIS LATE MEDWATCH REPORT: DURING A FEBRUARY 2019 INTERNAL REVIEW OF POST-MARKET SURVEILLANCE DATA FOR ALL APOLLO PRODUCTS, APOLLO EXECUTIVE MANAGEMENT WAS MADE AWARE THAT COMPLAINTS ASSOCIATED WITH THE CE MARKED ORBERA365 (12-MONTH INTRAGASTRIC BALLOON) WERE NOT BEING ASSESSED FOR REPORTABILITY CONSISTENT WITH THE REQUIREMENTS OF 21 CFR 803. CONSIDERATION FOR REPORTABILITY IN THE US WAS INCORRECTLY ESTABLISHED BASED ON THE DIFFERENCE IN THE INDICATIONS FOR USE (I.E. 12-MONTH PLACEMENT VS 6-MONTH PLACEMENT) VS CONSIDERATIONS FOR 'SIMILAR DEVICES' MARKETED BY APOLLO. THE ORBERA365 12-MONTH INTRAGASTRIC BALLOON IS A SIMILAR DEVICE' AS COMPARED TO THE ORBERA 6-MONTH BALLOON APPROVED VIA PMA P140008, AND AS SUCH, COMPLAINTS WILL BE ASSESSED FOR MDR REPORTABILITY GOING FORWARD. DEVICE EVALUATION SUMMARY: THE DEVICE WAS RETURNED TO APOLLO, AND A VISUAL INSPECTION WAS PERFORMED. IT WAS NOTED THAT BLACK PARTICLES WERE ON THE OUTER SURFACE OF THE BALLOON SHELL. A VALVE TEST WAS PERFORMED, AND WHEN DI WATER WAS INJECTED INTO THE VALVE, THE FLOW OF FLUID WAS CONTINUOUS AND UNOBSTRUCTED. AN AIR LEAK TEST WAS PERFORMED, AND LEAKAGE WAS NOTED FROM ONE OPENING ON THE ANTERIOR PORTION OF THE BALLOON SHELL. UNDER MICROSCOPIC ANALYSIS, THE OPENING WAS NOTED TO HAVE STRIATED EDGES, CONSISTENT WITH SURGICAL DAMAGE AND DEVICE REMOVAL ACTIVITIES. NO PARTICLES OR FOREIGN MATERIAL WAS NOTED IN THE VALVE CHANNEL. A REVIEW OF THE DEVICE LABELING NOTES THE FOLLOWING: WARNINGS AND PRECAUTIONS: THE RISK OF BALLOON DEFLATION AND INTESTINAL OBSTRUCTION (AND THEREFORE POSSIBLE DEATH RELATED TO INTESTINAL OBSTRUCTION) MAY BE HIGHER WHEN BALLOONS ARE LEFT IN PLACE LONGER THAN 12 MONTHS OR USED AT LARGER VOLUMES (GREATER THAN 700 CC). THE PHYSIOLOGICAL RESPONSE OF THE PATIENT TO THE PRESENCE OF THE ORBERA365¿ SYSTEM BALLOON MAY VARY DEPENDING UPON THE PATIENT'S GENERAL CONDITION AND THE LEVEL AND TYPE OF ACTIVITY. THE TYPES AND FREQUENCY OF ADMINISTRATION OF DRUGS OR DIET SUPPLEMENTS AND THE OVERALL DIET OF THE PATIENT MAY ALSO AFFECT THE RESPONSE. EACH PATIENT MUST BE MONITORED CLOSELY DURING THE ENTIRE TERM OF TREATMENT IN ORDER TO DETECT THE DEVELOPMENT OF POSSIBLE COMPLICATIONS. EACH PATIENT SHOULD BE INSTRUCTED REGARDING SYMPTOMS OF DEFLATION, GASTROINTESTINAL OBSTRUCTION, ACUTE PANCREATITIS, SPONTANEOUS INFLATION, ULCERATION AND OTHER COMPLICATIONS WHICH MIGHT OCCUR, AND SHOULD BE ADVISED TO CONTACT HIS/HER PHYSICIAN IMMEDIATELY UPON THE ONSET OF SUCH SYMPTOMS. COMPLICATIONS- POSSIBLE COMPLICATIONS OF THE USE OF THE ORBERA365¿ SYSTEM INCLUDE: BALLOON DEFLATION AND SUBSEQUENT REPLACEMENT.

Description of Event or Problem · 1

REPORTED AS: A PATIENT WITH THE ORBERA INTRAGASTRIC BALLOON HAD "LOSS OF LIQUID THROUGH VALVE." PATIENT HAD BLUE URINE. THE DEVICE WAS REMOVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
900624 ORBERA365 INTRAGASTRIC BALLOON SYSTEM INTRAGASTRIC BALLOON LTI APOLLO ENDOSURGERY, INC. B-50012 AF01994

Patients

Seq Age Sex Outcome Treatment
1 59 YR