ORBERA365 INTRAGASTRIC BALLOON SYSTEM
Report
- Report Number
- 3006722112-2019-00088
- Event Type
- Malfunction
- Date Received
- September 20, 2019
- Date of Event
- November 27, 2018
- Report Date
- November 30, 2018
- Manufacturer
- APOLLO ENDOSURGERY, INC.
- Product Code
- LTI
- PMA / PMN Number
- P140008
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- PHYSICIAN
Narratives
ADDITIONAL INFORMATION REGARDING THIS LATE MEDWATCH REPORT: DURING A FEBRUARY 2019 INTERNAL REVIEW OF POST-MARKET SURVEILLANCE DATA FOR ALL APOLLO PRODUCTS, APOLLO EXECUTIVE MANAGEMENT WAS MADE AWARE THAT COMPLAINTS ASSOCIATED WITH THE CE MARKED ORBERA365 (12-MONTH INTRAGASTRIC BALLOON) WERE NOT BEING ASSESSED FOR REPORTABILITY CONSISTENT WITH THE REQUIREMENTS OF 21 CFR 803. CONSIDERATION FOR REPORTABILITY IN THE US WAS INCORRECTLY ESTABLISHED BASED ON THE DIFFERENCE IN THE INDICATIONS FOR USE (I.E. 12-MONTH PLACEMENT VS 6-MONTH PLACEMENT) VS CONSIDERATIONS FOR 'SIMILAR DEVICES' MARKETED BY APOLLO. THE ORBERA365 12-MONTH INTRAGASTRIC BALLOON IS A [?]SIMILAR DEVICE' AS COMPARED TO THE ORBERA 6-MONTH BALLOON APPROVED VIA PMA P140008, AND AS SUCH, COMPLAINTS WILL BE ASSESSED FOR MDR REPORTABILITY GOING FORWARD. DEVICE EVALUATION SUMMARY: THE DEVICE WAS RETURNED TO APOLLO, AND A VISUAL INSPECTION WAS PERFORMED. THE BALLOON SHELL WAS NOTED TO BE DISCOLORED, AS IT WAS BLUE IN APPEARANCE. TRACE AMOUNTS OF BLUE FLUID WERE NOTED IN THE BALLOON. A VALVE TEST WAS PERFORMED, AND WHEN DI WATER WAS INJECTED INTO THE BALLOON VALVE, THE FLOW OF FLUID WAS CONTINUOUS AND UNOBSTRUCTED. AN AIR LEAK TEST WAS PERFORMED, AND LEAKAGE WAS NOTED FROM SEVERAL OPENINGS ON THE BALLOON SHELL. UNDER MICROSCOPIC ANALYSIS, FIFTEEN OPENINGS WERE NOTED ON THE BALLOON SHELL: NINE ON THE ANTERIOR PORTION OF THE SHELL, FIVE ON THE RADIUS OF THE SHELL, AND ONE ON THE POSTERIOR PORTION OF THE SHELL. ALL OPENINGS ON THE BALLOON SHELL WERE NOTED TO HAVE STRIATED EDGES, CONSISTENT WITH SURGICAL DAMAGE FROM DEVICE REMOVAL ACTIVITIES. BLACK PARTICLES WERE NOTED ON THE INNER SURFACE OF THE VALVE CHANNEL. A REVIEW OF THE DEVICE LABELING NOTES THE FOLLOWING: WARNINGS AND PRECAUTIONS: THE RISK OF BALLOON DEFLATION AND INTESTINAL OBSTRUCTION (AND THEREFORE POSSIBLE DEATH RELATED TO INTESTINAL OBSTRUCTION) MAY BE HIGHER WHEN BALLOONS ARE LEFT IN PLACE LONGER THAN 12 MONTHS OR USED AT LARGER VOLUMES (GREATER THAN 700 CC). THE PHYSIOLOGICAL RESPONSE OF THE PATIENT TO THE PRESENCE OF THE ORBERA365¿ SYSTEM BALLOON MAY VARY DEPENDING UPON THE PATIENT'S GENERAL CONDITION AND THE LEVEL AND TYPE OF ACTIVITY. THE TYPES AND FREQUENCY OF ADMINISTRATION OF DRUGS OR DIET SUPPLEMENTS AND THE OVERALL DIET OF THE PATIENT MAY ALSO AFFECT THE RESPONSE. EACH PATIENT MUST BE MONITORED CLOSELY DURING THE ENTIRE TERM OF TREATMENT IN ORDER TO DETECT THE DEVELOPMENT OF POSSIBLE COMPLICATIONS. EACH PATIENT SHOULD BE INSTRUCTED REGARDING SYMPTOMS OF DEFLATION, GASTROINTESTINAL OBSTRUCTION, ACUTE PANCREATITIS, SPONTANEOUS INFLATION, ULCERATION AND OTHER COMPLICATIONS WHICH MIGHT OCCUR, AND SHOULD BE ADVISED TO CONTACT HIS/HER PHYSICIAN IMMEDIATELY UPON THE ONSET OF SUCH SYMPTOMS. COMPLICATIONS- POSSIBLE COMPLICATIONS OF THE USE OF THE ORBERA365¿ SYSTEM INCLUDE: BALLOON DEFLATION AND SUBSEQUENT REPLACEMENT.
REPORTED AS: A PATIENT WITH THE ORBERA INTRAGASTRIC BALLOON HAD "BLUE URINE 8 HOURS AFTER THE IMPLANTATION." ADDITIONAL INFORMATION NOTED: THE BALLOON WAS IMPLANTED WITHOUT INCIDENTS OR LEAKS AND THE SAME DAY, 8 HOURS LATER, THE PATIENT CAME BACK AS [THEY] HAD OBSERVED BLUE URINE DUE TO THE METHILINE BLUE. A GASTROPOCY WAS CARRIED OUT WHICH SHOWED THE BALLOON TO BE MINIMALLY FULL WITH THE REST BLUE IN THE GASTRIC CAVITY, THEREFORE ACCORDING TO PROTOCOL, IT WAS DECIDED TO REMOVE THE BALLOON AND IMPLANT A REPLACEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 891020 | ORBERA365 INTRAGASTRIC BALLOON SYSTEM | INTRAGASTRIC BALLOON | LTI | APOLLO ENDOSURGERY, INC. | B-50012 | AF01892 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 32 YR |