FDA Adverse Event Other Summary report: N

PROGRAMMING SOFTWARE

MDR report key: 909835 · Received September 7, 2007

Report

Report Number
1644487-2007-01418
Event Type
Other
Date Received
September 7, 2007
Date of Event
August 10, 2007
Report Date
August 10, 2007
Manufacturer
CYBERONICS, INC.
Product Code
LYJ
PMA / PMN Number
P970003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

MFR #1644487-2007-01418. THE REPORTER INDICATED THAT A DELL HANDHELD WAS FROZEN IN THE WINDOWS SCREEN. A HARD RESET OF THE HANDHELD CORRECTED THE FROZEN SCREEN ISSUE. REPORTER FURTHER INDICATED THAT WHEN ATTEMPTING TO INTERROGATE A DEVICE, THE SCREEN WENT BACK TO THE WINDOWS SCREEN WITHOUT INDICATING A SUCCESSFUL INTERROGATION. THE NCP CYBERONICS 7.1 SCREEN WAS SELECTED AND THERE WAS A PROMPT REQUESTING "A NEW SET UP" INDICATED. AFTER SELECTING A NEW SET UP THERE WAS A COMMUNICATION ERROR; "ERROR 36 HAS OCCURRED". TROUBLESHOOTING DID NOT RESOLVE THE COMMUNICATION ERROR. THE DELL HANDHELD WAS RETURNED TO THE MANUFACTURER AND ANALYSIS WAS PERFORMED. PRODUCT ANALYSIS INDICATED THAT THE DATABASES CONTAINED 2 SEPARATE SETS OF DATA. THE MOST LIKELY CAUSE FOR THE 2 DIFFERENT SETS OF DATA CAN BE ASSOCIATED WITH A NEW DATABASE THAT WAS CREATED. THE ANALYSIS COULD NOT DETERMINE WHY A NEW DATABASE WAS CREATED BASED ON THE RETURNED FLASHCARD AND SOFTWARE. DURING THE ANALYSIS 2 CORRUPT ARCHIVE DATABASES WERE IDENTIFIED. THE MOST LIKELY CAUSE FOR THE DATABASE CORRUPTIONS IS ASSOCIATED WITH COMPATIBILITY ISSUES INVOLVING THE SANDISK FLASHCARD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PROGRAMMING SOFTWARE LYJ CYBERONICS, INC. MODEL 250 488813

Patients

Seq Age Sex Outcome Treatment
1 YR