PROGRAMMING SOFTWARE
Report
- Report Number
- 1644487-2007-01418
- Event Type
- Other
- Date Received
- September 7, 2007
- Date of Event
- August 10, 2007
- Report Date
- August 10, 2007
- Manufacturer
- CYBERONICS, INC.
- Product Code
- LYJ
- PMA / PMN Number
- P970003
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
MFR #1644487-2007-01418. THE REPORTER INDICATED THAT A DELL HANDHELD WAS FROZEN IN THE WINDOWS SCREEN. A HARD RESET OF THE HANDHELD CORRECTED THE FROZEN SCREEN ISSUE. REPORTER FURTHER INDICATED THAT WHEN ATTEMPTING TO INTERROGATE A DEVICE, THE SCREEN WENT BACK TO THE WINDOWS SCREEN WITHOUT INDICATING A SUCCESSFUL INTERROGATION. THE NCP CYBERONICS 7.1 SCREEN WAS SELECTED AND THERE WAS A PROMPT REQUESTING "A NEW SET UP" INDICATED. AFTER SELECTING A NEW SET UP THERE WAS A COMMUNICATION ERROR; "ERROR 36 HAS OCCURRED". TROUBLESHOOTING DID NOT RESOLVE THE COMMUNICATION ERROR. THE DELL HANDHELD WAS RETURNED TO THE MANUFACTURER AND ANALYSIS WAS PERFORMED. PRODUCT ANALYSIS INDICATED THAT THE DATABASES CONTAINED 2 SEPARATE SETS OF DATA. THE MOST LIKELY CAUSE FOR THE 2 DIFFERENT SETS OF DATA CAN BE ASSOCIATED WITH A NEW DATABASE THAT WAS CREATED. THE ANALYSIS COULD NOT DETERMINE WHY A NEW DATABASE WAS CREATED BASED ON THE RETURNED FLASHCARD AND SOFTWARE. DURING THE ANALYSIS 2 CORRUPT ARCHIVE DATABASES WERE IDENTIFIED. THE MOST LIKELY CAUSE FOR THE DATABASE CORRUPTIONS IS ASSOCIATED WITH COMPATIBILITY ISSUES INVOLVING THE SANDISK FLASHCARD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PROGRAMMING SOFTWARE | LYJ | CYBERONICS, INC. | MODEL 250 | 488813 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|