PULSE GEN MODEL 106
Report
- Report Number
- 1644487-2019-01827
- Event Type
- Malfunction
- Date Received
- September 20, 2019
- Date of Event
- July 17, 2015
- Report Date
- October 15, 2019
- Manufacturer
- CYBERONICS - HOUSTON
- Product Code
- LYJ
- UDI-DI
- 05425025750061
- PMA / PMN Number
- P970003
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
DESCRIBE EVENT OR PROBLEM, CORRECTED DATA: INITIAL REPORT INADVERTENTLY LEFT OUT INFORMATION REGARDING THE REVIEW OF THE DEVICE HISTORY RECORDS AND CONFIRMATION OF LASER ROUTING. (B)(4).
A REVIEW OF DEVICE HISTORY RECORDS REVEALED THAT THE GENERATOR PASSED QUALITY CONTROL INSPECTION PRIOR TO DISTRIBUTION. PER THE PRODUCT ANALYSIS, THE DEVICE WAS CONFIRMED AS LASER ROUTED.
IT WAS REPORTED THAT THE PATIENT UNDERWENT VNS GENERATOR REPLACEMENT SURGERY DUE TO BATTERY DEPLETION. THE EXPLANTED GENERATOR WAS RECEIVED BY THE MANUFACTURER. GENERATOR PRODUCT ANALYSIS WAS COMPLETED. THE REPORTED END OF SERVICE, OR EOS, CONDITION WAS DUPLICATED IN THE PRODUCT ANALYSIS, OR PA, LAB. DURING INTERROGATION, THE PULSE DISABLED AND EOS WARNINGS WERE SET. THE PULSE DISABLED BYTE WOULD NOT RESET AND, THEREFORE, SYSTEM DIAGNOSTICS AND THE FET COULD NOT BE PERFORMED. THE DATA IN THE DIAGACCUM CONSUMED MEMORY LOCATIONS REVEALED THAT 52.606% OF THE BATTERY HAD BEEN CONSUMED. WITH THE CASE REMOVED AND THE BATTERY STILL ATTACHED TO THE PRINTED CIRCUIT BOARD ASSEMBLY, OR PCBA, THE BATTERY MEASURED 1.824 V, CONFIRMING THE EOS CONDITION. CONTAMINATES WERE OBSERVED ON THE TRIMMED EDGE OF THE PCBA. THE BATTERY WAS REMOVED AND THE PCBA WAS SUBJECTED TO A POSTBURN ELECTRICAL TEST AND FAILED SEVERAL TESTS. WITH THE REMOVAL OF THE OBSERVED CONTAMINATES, THE RESULTS OF THE POSTBURN ELECTRICAL TEST WERE AS EXPECTED. THE OBSERVED CONTAMINATION SUGGESTED PROBABLE ELECTRICAL PATHS WERE ESTABLISHED BETWEEN THE COPPER EDGES ON THE TRIMMED EDGE OF THE PCBA, WHICH CONTRIBUTED TO THE SUPPLY CURRENT CONDITIONS AND RESULTED IN INCREASED OUT OF SPECIFICATION CURRENT CONSUMPTION FOR BOTH STANDBY AND PULSING MODES. THIS MAY HAVE BEEN A CONTRIBUTING FACTOR FOR THE REPORTED EOS. NO ADDITIONAL RELEVANT INFORMATION HAS BEEN RECEIVED TO DATE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 894342 | PULSE GEN MODEL 106 | GENERATOR | LYJ | CYBERONICS - HOUSTON | 106 | 203424 | 05425025750061 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 16 YR |