FDA Adverse Event Injury Summary report: N

MONAGHAN 225/SIMV VENTILATOR

MDR report key: 90947 · Received May 9, 1997

Report

Report Number
90947
Event Type
Injury
Date Received
May 9, 1997
Date of Event
August 29, 1996
Report Date
April 11, 1997
Manufacturer
MONAGHAN MEDICAL CORP.
Product Code
IKN
Product Problem
Yes
Report Source
User Facility report
Reporter Location
OK, US
Reporter Occupation
NURSE

Narratives

Description of Event or Problem · 1

PLACED VENT ON POSTOP PT. PT WAS ASSESSED, OBSERVED NO VENTILATION AND VENT WAS MAKING A LOUD SOUND, LIKE AIR ESCAPING. REPLACED VENT WITH ANOTHER MACHINE. FIOMED FINDINGS, FOUND PRESSURE LINE INSIDE VENTILATOR FROM THE MANIFOLD TO THE VENTILATOR PORT HAD COME LOOSE. CONNECTED PRESSURE HOSE TO CORRECT PORT AND RAN CHECK OUT PROCEDURE. ADJUSTED ASSIST SENSITIVITY CONTROL CHECK OUT 2/4 PER SERVICE MANUAL. PASSED. SETTINGS TIDAL VOL 800, RATE 12, 02 100%, PRESSURE 38.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MONAGHAN 225/SIMV VENTILATOR VENTILATOR IKN MONAGHAN MEDICAL CORP. 225-SIMV M4022

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Required Intervention