FDA Adverse Event
Injury
Summary report: N
MONAGHAN 225/SIMV VENTILATOR
MDR report key: 90947
·
Received May 9, 1997
Report
- Report Number
- 90947
- Event Type
- Injury
- Date Received
- May 9, 1997
- Date of Event
- August 29, 1996
- Report Date
- April 11, 1997
- Manufacturer
- MONAGHAN MEDICAL CORP.
- Product Code
- IKN
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- OK, US
- Reporter Occupation
- NURSE
Narratives
Description of Event or Problem · 1
PLACED VENT ON POSTOP PT. PT WAS ASSESSED, OBSERVED NO VENTILATION AND VENT WAS MAKING A LOUD SOUND, LIKE AIR ESCAPING. REPLACED VENT WITH ANOTHER MACHINE. FIOMED FINDINGS, FOUND PRESSURE LINE INSIDE VENTILATOR FROM THE MANIFOLD TO THE VENTILATOR PORT HAD COME LOOSE. CONNECTED PRESSURE HOSE TO CORRECT PORT AND RAN CHECK OUT PROCEDURE. ADJUSTED ASSIST SENSITIVITY CONTROL CHECK OUT 2/4 PER SERVICE MANUAL. PASSED. SETTINGS TIDAL VOL 800, RATE 12, 02 100%, PRESSURE 38.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MONAGHAN 225/SIMV VENTILATOR | VENTILATOR | IKN | MONAGHAN MEDICAL CORP. | 225-SIMV | M4022 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | Required Intervention |