FDA Adverse Event Malfunction Summary report: N

LIFEPAK CR PLUS DEFIBRILLATOR

MDR report key: 909195 · Received September 5, 2007

Report

Report Number
3015876-2007-00306
Event Type
Malfunction
Date Received
September 5, 2007
Report Date
August 13, 2007
Manufacturer
PHYSIO-CONTROL, INC.
Product Code
MKJ
PMA / PMN Number
K011144
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

PHYSIO-CONTROL, INC. EVALUATED THE DEVICE. THE ROOT CAUSE WAS DETERMINED TO BE DUE TO A FAILURE OF IC U10 ON THE ANALOG PCB ASSEMBLY. A REPLACEMENT DEVICE WAS PROVIDED TO THE CUSTOMER.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEFIBRILLATOR FAILED TO CHARGE APPROPRIATELY. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEPAK CR PLUS DEFIBRILLATOR MKJ PHYSIO-CONTROL, INC. LR CR PLUS NA

Patients

Seq Age Sex Outcome Treatment
1 NA YR