BIVONA 4.5 TTS HYPERFLEX FLEXTEND TRACHEOSTOMY TUBE
Report
- Report Number
- 3012307300-2019-04989
- Event Type
- Malfunction
- Date Received
- September 19, 2019
- Date of Event
- August 1, 2019
- Report Date
- October 23, 2019
- Manufacturer
- SMITHS MEDICAL ASD, INC.
- Product Code
- JOH
- UDI-DI
- 15019517056471
- PMA / PMN Number
- K923878
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
EVALUATION RESULTS: TWO CUSTOMIZED 4.5MM TRACHEOSTOMY TUBES (ID HYPERFLEX FLEXTENDS WITH FIXED CONNECTORS AND TTS CUFFS) WERE RETURNED FOR INVESTIGATION BECAUSE THE CUSTOMER WAS UNABLE TO INFLATE THE DEVICES. BOTH DEVICES WERE INFLATED USING 7CC'S OF AIR. BOTH DEVICES FULLY INFLATED WITH A SMALL AREA STUCK TO THE SHAFT, CAUSING THE CUFF TO APPEAR ASYMMETRICAL. THE CUFFS WERE DEFLATED AND GENTLY MASSAGED, AS SPECIFIED IN THE INSTRUCTIONS FOR USE (IFU). THE CUFFS WERE THEN INFLATED AGAIN WITH 7CC'S OF AIR AND THE CUFF INFLATED PROPERLY. BOTH DEVICES WERE LEAK TESTED IN A WATER BATH AND NO LEAKS WERE DETECTED. AFTER THE CUFF WAS MANIPULATED, AS STATED IN THE IFU, THEY INFLATED PROPERLY AND WERE SYMMETRICAL.
RELATED TO MFR NUMBER: 3012307300-2019-04990 (SAME PATIENT WITH TWO SAME DEVICE ISSUES).
INFORMATION WAS RECEIVED THAT A SMITHS MEDICAL BIVONA 4.5 TTS HYPERFLEX FLEXTEND TRACHEOSTOMY TUBE CUFF WAS NOT INFLATING PROPERLY WHILE IN USE WITH A PATIENT. THE REPORTER STATED THE PATIENT'S MOTHER WAS IN THE PROCESS OF A TRACHEOSTOMY TUBE CHANGE, BUT WAS UNABLE TO COMPLETE THE CHANGE OUT DUE TO THE EVENT. IT WAS ALSO REPORTED THE PATIENT HAD AN OLD TUBE IN THE HOUSE THAT WAS ABLE TO BE CLEANED AND USED UNTIL A NEW TUBE ARRIVED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 879548 | BIVONA 4.5 TTS HYPERFLEX FLEXTEND TRACHEOSTOMY TUBE | TUBE TRACHEOSTOMY AND TUBE CUFF | JOH | SMITHS MEDICAL ASD, INC. | XT16DS45NGF009S | DS025187 | 15019517056471 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |