FDA Adverse Event Malfunction Summary report: N

BIVONA 4.5 TTS HYPERFLEX FLEXTEND TRACHEOSTOMY TUBE

MDR report key: 9090327 · Received September 19, 2019

Report

Report Number
3012307300-2019-04990
Event Type
Malfunction
Date Received
September 19, 2019
Date of Event
August 1, 2019
Report Date
October 25, 2019
Manufacturer
SMITHS MEDICAL ASD, INC.
Product Code
JOH
UDI-DI
15019517056471
PMA / PMN Number
K923878
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE RESULTS OF THE PRODUCT EVALUATION (BELOW) WERE NOT INCLUDED IN THE INITIAL SUPPLEMENTAL REPORT. THIS CORRECTION REPORT NOW INCLUDES THE INVESTIGATION RESULTS OF THE RETURNED PRODUCT. EVALUATION RESULTS: TWO CUSTOMIZED 4.5 MM TRACHEOSTOMY TUBES (ID HYPERFLEX FLEXTENDS WITH FIXED CONNECTORS AND TTS CUFFS) WERE RETURNED FOR INVESTIGATION BECAUSE THE DEVICES WERE UNABLE TO INFLATE. BOTH DEVICES WERE INFLATED USING 7 CC'S OF AIR. BOTH DEVICES WERE FULLY INFLATED AND THE INVESTIGATOR NOTES A SMALL AREA STUCK TO THE SHAFT CAUSING THE CUFF TO APPEAR ASYMMETRICAL. THE CUFFS WERE THEN DEFLATED AND GENTLY MASSAGED, AS SPECIFIED IN THE INSTRUCTIONS FOR USE (IFU). THE CUFFS WERE INFLATED AGAIN WITH 7 CC'S OF AIR AND THE CUFFS INFLATED PROPERLY. BOTH DEVICES WERE LEAK TESTED IN A WATER BATH AND NO LEAKS WERE DETECTED.

Additional Manufacturer Narrative · 1

RELATED TO MFR NUMBER: 3012307300-2019-04989 (SAME PATIENT WITH TWO SAME DEVICE ISSUES).

Description of Event or Problem · 1

INFORMATION WAS RECEIVED THAT A SMITHS MEDICAL BIVONA 4.5 TTS HYPERFLEX FLEXTEND TRACHEOSTOMY TUBE CUFF WAS NOT INFLATING PROPERLY WHILE IN USE WITH A PATIENT. THE REPORTER STATED THE PATIENT'S MOTHER WAS IN THE PROCESS OF A TRACHEOSTOMY TUBE CHANGE, BUT WAS UNABLE TO COMPLETE THE CHANGE OUT DUE TO THE EVENT. IT WAS ALSO REPORTED THE PATIENT HAD AN OLD TUBE IN THE HOUSE THAT WAS ABLE TO BE CLEANED AND USED UNTIL A NEW TUBE ARRIVED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
879546 BIVONA 4.5 TTS HYPERFLEX FLEXTEND TRACHEOSTOMY TUBE TUBE TRACHEOSTOMY AND TUBE CUFF JOH SMITHS MEDICAL ASD, INC. XT16DS45NGF009S DS025187 15019517056471

Patients

Seq Age Sex Outcome Treatment
1