FDA Adverse Event Malfunction Summary report: N

MICROAIRE

MDR report key: 90876 · Received May 6, 1997

Report

Report Number
2020601-1997-00036
Event Type
Malfunction
Date Received
May 6, 1997
Date of Event
November 6, 1996
Report Date
May 5, 1997
Manufacturer
MICROAIRE SURGICAL INSTRUMENTS, INC.
Product Code
DZA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INSTRUMENT RECEIVED ON 5/9/97, TESTED PER THE ACCEPTANCE CRITERIA. INSTRUMENT OVERHEATED. TEAR DOWN OF INSTRUMENT SHOWED SPRING HAD CAUGHT ON RETAINER AND COILED AROUND SHAFT CAUSING FRICTION POINT. THIS PROBLEM HAS BEEN PREVIOUSLY INDENTIFIED AND HAS BEEN CORRECTED WITH REVISION TO THE COMPONENT AND ASSEMBLY INSTRUCTIONS. THE USER MANUAL OUTLINES MONITORING OF INSTRUMENT FOR OVERHEAT DURING USE AND COOLING INSTRUCTIONS.

Description of Event or Problem · 1

IT WAS REPORTED DURING A CALL TO THE HOSPITAL TO RETRIEVE THE LOANER INSTRUMENT THAT THE DRILL ALLEGEDLY BURNED THE LOWER RIGHT LIP OF THE PT. NO DEGREE OF BURN WAS NOTED IN THE PT RECORD. THE PT WAS TREATED WITH SALINE RINSES. NO OTHER INTERVENTION OR ADDITIONAL PROCEDURE WAS USED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MICROAIRE HIGH SPEED DRILL DZA MICROAIRE SURGICAL INSTRUMENTS, INC. 2910-100 4C418

Patients

Seq Age Sex Outcome Treatment
1 32 YR