FDA Adverse Event Malfunction Summary report: N

UNKNOWN

MDR report key: 9084853 · Received September 18, 2019

Report

Report Number
2249723-2019-01497
Event Type
Malfunction
Date Received
September 18, 2019
Date of Event
August 22, 2019
Report Date
November 8, 2019
Manufacturer
DATASCOPE CORP. - MAHWAH
Product Code
DSP
PMA / PMN Number
UNKNOWN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE SUSPECTED FAULTY OOB COMPRESSOR WAS RETURNED TO GETINGE¿S NATIONAL REPAIR CENTER (NRC) FOR EVALUATION. A SENIOR REPAIR TECHNICIAN INSPECTED THE COMPRESSOR AND FOUND THE JACKET OF THE CABLE HARNESS COMING OUT OF THE MOTOR RIPPED OPEN WITH THE WIRES EXPOSED. THE SENIOR REPAIR TECHNICIAN ALSO FOUND BROKEN RED WIRE(MOTOR A) NICKED INSULATION WITH BARE WIRE SHOWING ON GREEN WIRE(HALL -) AND THE BROWN WIRE(HALL A). THE NATIONAL REPAIR CENTER VERIFIED THE REPORTED FAILURE OF "NEW COMPRESSOR WAS FOUND IN DAMAGED CONDITION WHEN THE BOX WAS OPENED". SCRAPPED AND RETAINING THE PUMP ASSEMBLY IN THE NRC PER PROCEDURE.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE CABLE OF A NEW COMPRESSOR WAS FOUND TO BE IN DAMAGED CONDITION WHEN THE BOX WAS OPENED BY A GETINGE REPRESENTATIVE. THIS IS AN OUT-OF-BOX (OOB) FAILURE. THERE WAS NO PATIENT INVOLVEMENT; THUS NO ADVERSE EVENT WAS REPORTED.

Additional Manufacturer Narrative · 1

A PRODUCTION DEVICE HISTORY RECORD (DHR) REVIEW IS NOT REQUIRED AS THIS COMPLAINT IS FOR AN OUT-OF-BOX PART FAILURE, AND THERE WAS NO ALLEGED MALFUNCTION OF AN INTRA-AORTIC BALLOON PUMP (IABP). A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION IS MADE AVAILABLE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CABLE OF A NEW COMPRESSOR WAS FOUND TO BE IN DAMAGED CONDITION WHEN THE BOX WAS OPENED BY A GETINGE REPRESENTATIVE. THIS IS AN OUT-OF-BOX (OOB) FAILURE. THERE WAS NO PATIENT INVOLVEMENT; THUS NO ADVERSE EVENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
876033 UNKNOWN SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL DSP DATASCOPE CORP. - MAHWAH N/A N/A