ENDOPATH ENDOSCOPIC LINEAR CUTTER
Report
- Report Number
- 1527736-1997-00883
- Event Type
- Malfunction
- Date Received
- May 8, 1997
- Date of Event
- January 1, 1997
- Report Date
- May 8, 1997
- Manufacturer
- ETHICON ENDO-SURGERY, INC.
- Product Code
- GAG
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
FACILITY EXPERIENCED AN EVENT WITH THE ENDOPATH ENDOSCOPIC LINEAR CUTTER WHILE PERFORMING A L.A.V.H.. THE PRODUCT COMPLAINT ANALYSIS TEAM HAS COMPLETED ITS INVESTIGATION OF THE DEVICE WHICH WAS RETURNED TO CO WITH PRODUCT INQUIRY #972231. THE RESULTS OF THE INVESTIGATION CONDUCTED BY THE APPROPRIATE ENGINEERS AND TECHNICIANS ARE LISTED BELOW. VISUAL INSPECTIONS & RESULTS: LOCKOUT POSITION, FIRED; CARTRIDGE CONDITION, FULLY FIRED; CARTRIDGE RETURN BATCH NUMBER, J00C0E; AND INSTRUMENT NUMBER, 0116. FUNCTIONAL TESTS & RESULTS: CONDITION OF FIRING TRIGGER LOCKOUT, CONDITION OF PINION GEAR, CONDITION OF SHORT RACK, AND CONDITION OF YOKE, GOOD; AND TEST CARTRIDGE BATCH #, J00K0W. ANALYSIS CONCLUSION: BASED UPON THE INQUIRY INFO RECEIVED, THE VISUAL EXAMINATION, AND THE FUNCTIONAL TESTING, NO CONCLUSION COULD BE REACHED AS TO WHY THE INSTRUMENT REPORTEDLY "SNAPPED INSIDE THE PATIENT" DURING SURGERY. THE INSTRUMENT WAS RETURNED IN GOOD PHYSICAL CONDITION. THE INSTRUMENT WAS CYCLED, FIRED, CUT, AND FORMED THE STAPLES WITHIN DESIGN SPECIFICATION. THE INSTRUMENT WAS DISASSEMBLED TO EXAMINE THE INTERNAL COMPONENTS AND NO DEFORMATIONS COULD BE IDENTIFIED. IT WAS CONCLUDED THAT THE INSTRUMENT WAS FULLY FUNCTIONAL AND CONFORMING TO DESIGN SPECIFICATIONS. THE EXPERIENCE THE SURGEON REPORTED COULD NOT BE REPEATED. EACH INSTRUMENT IS EVALUATRED DURING THE ASSEMBLY PROCESS TO ENSURE IT FUNCTIONS PROPERLY. CO STRIVES TO UNDERSTAND EACH INCIDENT AS IT OCCURS IN ORDER TO CONTINUOUSLY IMPROVE THE PRODUCTS.
DURING A L.A.V.H. PROCEDURE, IT WAS REPORTED BY THE AFFILIATE THAT THE DEVICE SNAPPED INSIDE THE PT. THERE WAS NO PT CONSEQUENCE. A SECOND DEVICE WAS USED TO COMPLETE THE CASE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENDOPATH ENDOSCOPIC LINEAR CUTTER | ENDOSCOPIC LINEAR CUTTER | GAG | ETHICON ENDO-SURGERY, INC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | Other |