ZOLL IVTM THERMOGARD XP
Report
- Report Number
- 3010617000-2019-00838
- Event Type
- Malfunction
- Date Received
- September 17, 2019
- Date of Event
- August 25, 2019
- Report Date
- October 11, 2019
- Manufacturer
- ZOLL CIRCULATION
- Product Code
- NCX
- UDI-DI
- 00849111075022
- PMA / PMN Number
- K072234
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THERMOGARD SYSTEM WAS RECEIVED AT ZOLL SAN JOSE FOR FURTHER EVALUATION. THE REPORTED COMPLAINT OF FLICKERING DISPLAY SCREEN AND "M ID:05" (POST EXT RAM) ERROR MESSAGE ON THE THERMOGARD XP IVTM SYSTEM (SERIAL # (B)(4) WERE CONFIRMED DURING FUNCTIONAL TEST. THE ROOT CAUSE WAS DUE TO DAMAGED PROCESSOR PCA. THE PROCESSOR PCA WAS REPLACED TO REMEDY THE FLICKERING DISPLAY SCREEN AND "M ID:05" (POST EXT RAM). ARCHIVE COULD NOT BE RECOVERED AND REVIEW DUE TO DAMAGED PROCESSOR PCA. AFTER PART REPLACEMENT, THE SYSTEM WAS RE-EVALUATED AND PASSED ALL THE FUNCTIONAL TESTS WITHOUT ANY FAULT OR ERROR. THERMOGARD XP IVTM SYSTEM IS READY FOR IVTM THERAPY USE. HISTORICAL COMPLAINTS WERE REVIEWED FOR SERVICE INFORMATION RELATED TO THE REPORTED COMPLAINT AND THERE WERE NO SIMILAR COMPLAINTS REPORTED FOR THERMOGARD XP IVTM SYSTEM WITH SERIAL NUMBER (B)(4).
THE REPORTED COMPLAINT OF FLICKERING DISPLAY SCREEN AND "M ID:05" (POST EXT RAM) ERROR MESSAGE ON THE THERMOGARD XP IVTM SYSTEM (SERIAL (B)(4)) WERE CONFIRMED DURING FUNCTIONAL TEST. THE ROOT CAUSE WAS DUE TO DAMAGED DISPLAY HEAD. DURING VISUAL INSPECTION, NO PHYSICAL DAMAGE WAS OBSERVED ON THE CONSOLE. THE INITIAL FUNCTIONAL TESTING FAILED DUE TO FLICKERING SCREEN AND "M ID:05" (POST EXT RAM) ERROR MESSAGE. THERMOGARD SYSTEM WILL BE SENT TO ZOLL SAN JOSE FOR FURTHER EVALUATION. HISTORICAL COMPLAINTS WERE REVIEWED FOR SERVICE INFORMATION RELATED TO THE REPORTED COMPLAINT AND THERE WERE NO SIMILAR COMPLAINTS REPORTED FOR THERMOGARD XP IVTM SYSTEM WITH SERIAL NUMBER (B)(4).
CUSTOMER REPORTED THAT DURING WORKSHOP, THE DISPLAY SCREEN OF THERMOGARD XP IVTM SYSTEM (SERIAL (B)(4)) WAS FLICKERING. WHEN AZM TS SWITCHED THE DISPLAY HEAD WITH A KNOWN GOOD DISPLAY HEAD AT THE CUSTOMER SITE, NO PROBLEM OCCURRED. THEN, AZM TS BROUGHT BACK THE REMOVED SUSPECTED DISPLAY HEAD TO THE OFFICE AND ATTACHED TO ANOTHER TGXP CONSOLE (AZM'S DEVICE). THE REPORTED FLICKERING ISSUE OCCURRED AND "M ID:05" (POST EXT RAM) DISPLAYED ON THE SCREEN. NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 871079 | ZOLL IVTM THERMOGARD XP | THERMAL REGULATING SYSTEM | NCX | ZOLL CIRCULATION | 8700-0650 | 00849111075022 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |