FDA Adverse Event Malfunction Summary report: N

BD SYRINGE LUER-LOK TIP

MDR report key: 9070198 · Received September 16, 2019

Report

Report Number
1213809-2019-00929
Event Type
Malfunction
Date Received
September 16, 2019
Date of Event
August 28, 2019
Report Date
September 27, 2019
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
FMF
UDI-DI
30382903096283
PMA / PMN Number
K941562
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: ONE 1ML SYRINGE IN AN OPENED BLISTER PACK FROM BATCH 9112904 (P/N 309628) WAS RECEIVED AND EVALUATED. IT WAS OBSERVED THE STOPPER WAS NOT CONNECTED TO THE PLUNGER ROD AND WAS REJECTABLE PER PRODUCT SPECIFICATION. FURTHERMORE, A DEVICE HISTORY RECORD REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION OF THE PROVIDED LOT NUMBER THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. POTENTIAL ROOT CAUSE FOR THE STOPPER SEPARATION DEFECT IS ASSOCIATED WITH THE ASSEMBLY PROCESS.

Description of Event or Problem · 0

MATERIAL NO.: 309628. BATCH: 9112904. IT WAS REPORTED THAT BEFORE USE OF THE BD SYRINGE LUER-LOK¿ TIP THERE WAS STOPPER SEPARATION FROM PLUNGER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: IT WAS REPORTED THAT THERE WAS STOPPER SEPARATION FROM PLUNGER.

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

MATERIAL NO.: 309628. BATCH: 9112904. IT WAS REPORTED THAT BEFORE USE OF THE BD SYRINGE LUER-LOK¿ TIP THERE WAS STOPPER SEPARATION FROM PLUNGER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: IT WAS REPORTED THAT THERE WAS STOPPER SEPARATION FROM PLUNGER..

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
866842 BD SYRINGE LUER-LOK TIP PISTON SYRINGE FMF BECTON DICKINSON MEDICAL SYSTEMS 9112904 30382903096283

Patients

Seq Age Sex Outcome Treatment
1 Other