FDA Adverse Event Other Summary report: N

I-STAT CTNI CARTRIDGE

MDR report key: 906955 · Received August 30, 2007

Report

Report Number
2245578-2007-00028
Event Type
Other
Date Received
August 30, 2007
Date of Event
July 12, 2007
Report Date
August 30, 2007
Manufacturer
ABBOTT POINT OF CARE INC.
Product Code
MMI
PMA / PMN Number
K031739
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

IN 2007, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER WHO REPORTED THAT AN I-STAT CARDIAC TROPONIN I CARTRIDGE YIELDED A RESULT OF 0.27 NG/ML OR A PT AT 21:45 ABOUT 3 WEEKS BEFORE. THE PT WAS TRANSFERRED TO ROYAL MELBOURNE HOSPITAL WHERE A SAMPLE GENERATED A TROPONIN I RESULT OF <0.01 NG/ML USING AN UNK METHOD. THE I-STAT RESULT OF 0.27 NG/ML CONTRIBUTED TO THE DECISION TO PERFORM AND ANGIOGRAM ON THE PT. TEST RESULTS FROM THE ANGIOGRAM WERE NEGATIVE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 I-STAT CTNI CARTRIDGE IVD MMI ABBOTT POINT OF CARE INC. NA U07105

Patients

Seq Age Sex Outcome Treatment
1 YR Other