FDA Adverse Event Summary report: N

QUICK CHECK ONE

MDR report key: 90586 · Received May 9, 1997

Report

Report Number
2027242-1997-90151
Date Received
May 9, 1997
Date of Event
April 9, 1997
Report Date
April 10, 1997
Manufacturer
DIAGNOSTIC SOLUTIONS, INC.
Product Code
CFR
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
PATIENT

Narratives

Description of Event or Problem · 1

THE PATIENT REPORTED OUT OF RANGE NORMAL CONTROL SOLUTION TEST RESULTS WITH LOT 16089A. ON 4/9/97, THE PATIENT OPENED THE FIRST BOTTLE FROM THE BOX AND OBTAINED A NORMAL CONTROL SOLUTION TEST RESULT OF "190 OR SO" MG/DL VERSUS A NORMAL CONTROL SOLUTION RANGE OF 111-167 MG/DL (SOL. LOT UNK, EXP. 9/98). THE PATIENT SHOOK THE CONTROL SOLUTION BEFORE HE APPLIED THE SAMPLE. THE CONTROL SOLUTION WAS OPENED "2 OR MORE" MONTHS AGO. WITH THE CO'S REP, THE PATIENT OBTAINED TWO NORMAL CONTROL SOLUTION TEST RESULTS OF 161 AND 156 MG/DL. THE PATIENT'S METER WAS SET TO THE APPROPRIATE CODE. THE DESICCANT IS IN THE FIRST BOTTLE FROM THE BOX. THE PATIENT HAS NOT OPENED THE SECOND BOTTLE FROM THE SAME BOX. THE PATIENT STORES THE TEST STRIPS IN THE KITCHEN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 QUICK CHECK ONE BLOOD GLUCOSE TEST STRIPS FOR USE WITH ONE TOUCH METERS CFR DIAGNOSTIC SOLUTIONS, INC. * 16089A

Patients

Seq Age Sex Outcome Treatment
1 *